FW_F112P_v1901_02_EN.docx
Page 25
4.4.6.
TERMINAL CONNECTORS
Refer to Appendix A: Technical Specification
Fig. 8: Overview of terminal connectors - Standard configuration and options
4.4.7.
SENSOR SUPPLY
For type PB/PC; PX; AP:
There is no real sensor supply out available. Only a limited power supply is available. This power
supply MAY NOT be used to supply the flowmeters electronics, converters etc. as it will not provide
adequate sustained power ! All energy used by the flowmeters pick-up will directly influence the
battery life-time. It is strongly advised to use a "zero power" pickup such as a coil or reed-switch
when operating without external power. It is possible to use some low power NPN or PNP output
signals, but the battery life time will be significantly reduced (consult your distributor). The sensor
supply is fixed: 1.2V DC or 3V DC (set by the firmware).
For type PD; PF; PM:
It is possible to supply the sensor with different voltages. You can set the voltage with the switches.
Internal power is only applicable for low power sensors (Coil, Reed). External power is only available
when the main external power supply is connected.
The sensor supply voltage is selectable: 1.2; 3; 8.2; 12 or 24V DC.