FW_F112A_v1702_02_EN
Page 38
Terminal connections
Type:
Removable plug-in terminal strip. Wire max. 1.5mm
2
and 2.5mm
2
Data protection
Type
EEPROM backup of all setting. Backup of running totals every minute.
Data retention at least 10 years.
Password
Configuration settings can be password protected.
Hazardous area
Intrinsically safe
Type XI
ATEX approval :
II 1 G Ex ia IIB/IIC T4 Ga
II 1 D Ex ia IIIC T100°C Da
IECEx approval :
Ex ia IIB/IIC T4 Ga
Ex ia IIIC T100°C Da
Explosion proof
Type XD/XF
ATEX approval ref: II 2 EEx d IIB T5. Weight appr. 15kg.
Dimensions of enclosure: 350 x 250 x 200mm (13.7” x 9.9” x 7.9”) LxHxD.
Directives and
Standards
EMC
EN 61326-1; FCC 47 CFR part 15
LVD
EN/IEC 61010-1
ATEX / IECEx
EN/IEC 60079-0; EN/IEC 60079-11
RoHS
EN 50581
IP & NEMA
EN 60529; NEMA 250
Flowmeter
Type A
(0)4-20mA - with signal calibration feature at any current within the range.
Accuracy Resolution: 16 bit.. Error < 0.025mA / ±0.125% FS. Low level cut-off programmable.
Span 0.000010 - 9999999 with variable decimal position.
Update time Four times a second.
Voltage drop 2.5 Volt
Relationship Linear and square root calculation.
Note External power to sensor is required; e.g. type PD.
Linearization 15 positions with interpolation function;
Meter-Factor 0.000001 - 9.999999 versus Percentage (0)4-20mA.
15B
Reset
Type IB
Make contact - external reset totalizer. Additional functionality to lock the RESET button of the
keyboard (as long as this contact is being made).
Duration Minimum 100mSec. to reset Total.
INPUTS