Access Array System for the Ion Torrent PGM Sequencing System: User Guide
Chapter 3: Target-Specific Primer Validation for 4-Primer Amplicon Tagging on the LP 48.48 IFC
Check PCR Products on the Agilent 2100 Bioanalyzer
21
Check PCR Products on the Agilent 2100 Bioanalyzer
Use the Agilent DNA 1000 chips from the Agilent DNA 1000 Kit to check
1
μ
L of PCR product from
each of the PCR reactions described above. Follow the
Agilent DNA 1000 Kit Guide
for details.
On-target products should account for a minimum of 50% of the total yield (by mass)
produced for a particular primer pair. If on-target products comprise <50% of total yield,
primers may have to be redesigned.
NOTE
The product size should be the sum of the target region and the length of the
Access Array Barcode 1 primers for Ion Torrent (103 bp).
Figure 1. Example of 48 primer pairs checked on an Agilent chip. The red-labeled PCR reactions
generated multiple or incorrectly sized primers, which need to be redesigned and validated before
continuing.