Adjusting Clamping Force
10.
After clamping, read and record the force on the gauge.
11.
Place the power switch in the OFF position, wait about 5 seconds and return
it to the ON position.
12.
Press the UP button.
13.
If the force is at the lower limit or less, change the adjustment switch to the
next lowest number and go to step #9.
Note:
If the force is at the “upper limit” or greater, change the
adjustment switch to the next highest number and go to
step #9. If the force falls within the target, go to the next
step.
14.
Press the power OFF switch.
15.
Remove the gauge and spacer.
16.
Unplug the power cord from the electrical outlet.
17.
Place the wire back on the upper left terminal on the shake relay.
18.
Close the control box and insert the screws.
19.
Plug the power cord into the electrical outlet.
At this point, the mixer is ready for normal operation.
Fluid Management Customer Service 1.800.462.2466
Harbil 5G Mixer | 41