4. Installation
The battery powered flow monitor is available just in one packaging for compact field
version, panel or wall installation. The compact field version is mounted on top of the
sensor using the compact mounting kit (F9.KC1), the panel version is installed using the
panel mounting kit (F9.KP1), while the wall mounting version is fixed with the wall
mounting kit (F9.KW1).The mounting kits can be ordered directly connected to the monitor
or separately and then simply installed on it.
4.1. Panel Installation
The panel mounting version consists of the monitor and the mounting bracket kit
F9.KP1 with gasket for IP65 watertight panel installation. The monitor perfectly fits into a
standard ¼ DIN panel cutout.
1. Cut out the panel: the F9.20
requires a panel opening of
90,5 x 90,5 mm (3,563” x
3,563”). ¼ DIN punches are
recommended, alternatively a
jigsaw or another cutting tool
may be used.
2. Recommended
minimum
clearance between panel
cutouts is 28 mm (1.1”) as
illustrated.
3. Place gasket on the instrument
and install into the panel. Be sure
the panel gasket is properly
seated against the panel and
around the instrument case.
4. Slide the metal bracket over the
back of the instrument. Press the
bracket against the inside of the
panel to perfectly fix the
instrument in place.
To REMOVE: press the clips outward while pulling the bracket away from the
instrument panel. Do not allow the instrument to fall out of the panel opening: it
may be helpful to secure the instrument temporarily with tape from front.
F9.20 Battery Powered Flow Monitor
7