F9.20 Battery Powered Flow Monitor
13
6. View Level
•
During normal operation, the flow monitor is in View Level displaying all measured
values.
•
If the flow monitor is in a different level and no activity occurs for more than 3
minutes, it will return to View Level.
•
To select the item you want displayed, press UP or DOWN arrows.
•
Changing display indication does not affect or interrupt instrument operation
and calculation.
Description
Flow rate and resettable totalizer value
Resettable totalizer and flow rate value
Permanent totalizer value
Flow rate value
Resettable totalizer value. Press the RIGHT arrow key
to reset. If locked, you will need to enter the password
first. Lock or unlock the total reset in Option Menu (refer
to
section 8.4.8. Restot PWD
for details)
7. Menu Directory Level
Access to this level can be free or password protected. Entering the correct password
allows direct access to next levels and to all editable items in all menus, until a return to
View Level (refer to
section 8.4.7. Menu PWD
to select password protected access).
Four different menus are available to fully set-up the F9.00 flow monitor and transmitter.
These menus are separated in two different Menu Directories.
In terms of getting started and making measurements, Calibration Menu is the most
important menu in the F9.00 and it is the only one included in the first Menu Directory.
Output Menu, Simulation Menu and Option Menu are included together in the second
Menu Directory.