Flowserve.com 13
PUIOM000649-00-UIM (EN/AQ)
August 2022
SEM-100SS User Instructions
7. Press
“UP”
and
“DOWN”
arrow keys to enter the amount of fluid that
was pumped in the desired units.
Note:
The display will return to the Current Total screen and read the
previous total.
Note:
“Err” will be displayed if too few or too many counts have been
received during the calibration. A volume between 0.5 to 20 units must
be recorded for a valid calibration. If “Err” is displayed, no calibration
change has taken place.
CAL-3 Calibration:
1. Press
“CAL-1”
to enter calibration mode.
2. Press
“CAL-3”
. The current CAL-3 number will show.
3. Press the
“UP”
and
“DOWN”
arrow keys to adjust to a known
CAL-3 number.
4. Press
“CAL-1”
to enter the new number and complete the calibration
process.
Note:
Use of the CAL-3 function will override any previous calibrations.
Note:
A typical “CAL-3” number for water is 600. Recording the
“CAL-3” numbers for specific fluids after calibration, and resetting the
meter to those numbers when the fluid is pumped again is a convenient
method to quickly and accurately calibrate the meter.