23
Basic Settings
– continued –
Establishing measuring range and actual value output
– continued –
Code switch
Measuring range
(µS/cm at 25 °C)
Current output mA = µS/cm
1
2
3
4 mA is
equivalent to
20 mA is
equivalent to
OFF
OFF
OFF
100
3000
100
3000
ON
OFF
OFF
5000
5000
OFF
ON
OFF
7000
7000
Factory setting
ON
ON
OFF
10000
10000
ON
ON
ON
12000
12000
LRGT 16-2
When the supply voltage is applied (during commissioning) the first output will be 4 mA, then the output
current increases until it reaches the actual value.