Rev A1   

MN300480 

7 of 24 

GETTING STARTED (continued) 

Step Five 

USB®  Fob  Interface:   

EchoSonic®  II  communicates  with  WebCal®  through  a  USB®  interface  called  a  Fob.  

Before plugging your Fob into your computer’s USB® port, be sure that you have installed WebCal® on your 
computer. 
 
Connect  the  red,  green,  white  and  black  wires  from  EchoSonic®  II  into  the  correct  terminals  on  the  Fob.  
Tighten the screws on the terminals and plug your Fob into the USB® port of your computer.   
 

 

 

Wiring identical for all series – Use only the Red, Black, Green and White wires. 

 

The  maximum  cable  distance  between  the  computer  and  EchoSonic®  II  is  15’.    This  only  applies  when 
configuring the EchoSonic® II. 

 

Once  EchoSonic®  II  is  configured  and  prior  to  installation,  isolate  the  white  and  green  wires  from  active 
power to prevent a short of the configuration circuit 

 

Note:

 When using the Fob, do not add VDC power.  The Fob, when connected to the computer, will provide 

the required power to the EchoSonic® II.  

*

 

For  complete  information  on  the  WebCal®  software,  please  refer  to  the  WebCal®  manual  located  at 

www.flowline.com 

 

Note:  

EchoSonic

®

 II ships with the LI99‐2001 Fob (white in color).  The LI99‐1001 Fob (black in color) 

can be 

used 

with EchoSonic

®

 II (LU23, LU28 & LU29 series) to configure the sensor with WebCal

®

.     

 
 

 

 

Содержание EchoSonic II LU27 Series

Страница 1: ...0480 1 of 24 EchoSonic II Ultrasonic Level Transmitter LU23 LU27 LU28 LU29 Series Manual NEMA 4X Enclosure Flowline Inc 10500 Humbolt Street Los Alamitos CA 90720 Tel 562 598 3015 Fax 562 431 8507 www flowline com ...

Страница 2: ...mA to 20 4 mA Adjustable start up condition Empty Mid 12 mA Full Over range 22 mA Table of Contents Specifications 3 Dimensions 3 About this manual 4 Components 5 Getting Started 6 USB Fob Interface 7 Four Steps for Using WebCal 8 Step 1 Sensor Configuration 8 Step 2 Tank Shape Selection 9 Step 3 Tank Levels Configuration 9 Step 4 Write to Unit 10 Wiring 11 Wiring Diagram 11 Wiring Connections 11 ...

Страница 3: ...ail safe 4 mA 20 mA 21 mA 22 mA or hold last Process temp F 4 to 140 C 20 to 60 Temp comp Automatic Ambient temp F 31 to 140 C 35 to 60 Pressure MWP 30 PSI Enclosure type NEMA 4X IP65 Encl hardware Brass stainless Enclosure vent Water tight membrane Conduit entrance Dual NPT Encl material Polycarbonate Trans material PVDF Process mount LU27 1 NPT 1 G LU23 28 29 2 NPT 2 G Mount gasket Viton Classif...

Страница 4: ...mpatible with the application media Enclosure While the transmitter housing is liquid resistant the EchoSonic II is not designed to be operational when immersed It should be mounted in such a way that the enclosure and transducer do not come into contact with the application media under normal operational conditions The enclosure has a flip cover with dual 1 2 NPT female conduit ports and an inter...

Страница 5: ... 200128 for LU27 series o Part 220129 for LU23 LU28 or LU29 series USB Fob o Part LI99 2001 Quick Start Guide P N Max Range Dead Band Thread Fob LU23 40 18 0 5 5m 8 20cm 2 NPT No Fob LU23 41 Fob Included LU23 50 2 G No Fob LU23 51 Fob Included LU27 40 9 8 3m 4 10cm 1 NPT No Fob LU27 41 Fob Included LU27 50 1 G No Fob LU27 51 Fob Included LU28 40 26 2 8m 8 20cm 2 NPT No Fob LU28 41 Fob Included LU2...

Страница 6: ...stall WebCal prior to plugging in the USB Fob Please go to www flowline com webcal software and select your language version WebCal System Requirements Windows XP Vista 7 8 10 32 or 64 bit system 1 USB 2 0 port 10 mB hard drive space 256 mB RAM Internet connection For complete information on the WebCal software please refer to the WebCal manual located at www flowline com webca software ...

Страница 7: ...hite wires The maximum cable distance between the computer and EchoSonic II is 15 This only applies when configuring the EchoSonic II Once EchoSonic II is configured and prior to installation isolate the white and green wires from active power to prevent a short of the configuration circuit Note When using the Fob do not add VDC power The Fob when connected to the computer will provide the require...

Страница 8: ...ebCal software please refer to the WebCal manual located at www flowline com webcal software Configuring EchoSonic II with WebCal 1 Sensor Configuration a Configures the Loop Fail Safe Output at Empty and Startup condition for the sensor 2 Tank Shape Selection a Defines the shape of the tank as well as the dimensional information for the tank with respect to the sensor s location on the tank 3 Tan...

Страница 9: ...he tank as well as the dimensional information for the tank with respect to the sensor s location on the tank Tank Level Configuration The Height Units Sensor Height Fill Height and Capacity were all calculated in the previous Dimensional Entry window To adjust these settings click on Select Tank Shapes ...

Страница 10: ...ploads configuration into the sensor provides a custom wiring diagram specific to the signal output and or relay configuration and saves the configuration file to your hard drive For complete information on the WebCal software please refer to the WebCal manual located at www flowline com webcal php ...

Страница 11: ...y p e EchoSonic g diagram c II and you d R and ower supply wires can b wire howeve White W T Cal and sho These wires pplied from a cable distanc e or green te power supply M II has bee is shown a ur applicatio d Black BLK or to a 4 2 be extended er do not ext T and Gree ould not be s should not any source o ce between rminals or w y MN300480 en configure bove Flow on s compon K leads a 20 mA loop ...

Страница 12: ...al spikes by isolating the power whenever possible The sensor materials must be chemically compatible with the liquids to be measured Design a fail safe system for possible sensor and or power failure Never use the sensor in environments classified as hazardous Voltage Output EchoSonic II can be used as a 0 to 5 or 0 to 10 VDC output device A resistor will need to be added to the circuit to enable...

Страница 13: ...low is a quick review of wiring the EchoSonic II to common display controllers and PLC s DataView LI55 Series Level Controller Commander LI90 Series Multi Tank Level Controller DataLoop LI23 Series Level Indicator without the backlight DataLoop LI23 Series Level Indicator with the backlight ...

Страница 14: ... WIRING continued Step Six Wiring to Display Controllers PLC s continued DataPoint LC52 Series Level Controller JWA mode Factory Setting DataPoint LC52 Series Level Controller JWB mode Generic Loop Powered Display Generic PLC ...

Страница 15: ...a vacuum 6 Avoid mounting in the center of a dome top tank 7 In cone bottom tank position the sensor over the deepest part of the tank Installation in existing fittings If the existing fitting is larger than the threads of the EchoSonic II select a reducer bushing such as the LM52 1400 2 thread x 1 thread or LM52 2400 3 thread x 2 thread LM52 1400 LM52 2400 Do not install at angle relative to the ...

Страница 16: ... Slip x Thread LM52 2890 Tank Adapter w 2 x 1 Reducer Bushing LM52 2890 w LM52 1800 Tank Adapter Thread x Thread Do not use thread x thread 2 Riser Installations with tall narrow risers can impede the acoustic signal a Core Out Concrete Applications where tank with a concrete ceiling that has been cored out can also be considered as a riser type application In these applications follow a 2 1 ratio...

Страница 17: ...d a 2 to 1 reducer bushing to complete the installation 2 Flange w thread out of plane LM52 1850 2 Flange w thread in plane 2 Flange w Reducer Bushing w LM52 1800 Do not use thread in plane 4 Side Mount Bracket For installations in open tanks and sumps use the LM50 series side mount bracket a For the LU27 series order the LM50 1001 1 which includes a 2 x 1 Reducer Bushing b For the LU23 LU28 LU29 ...

Страница 18: ... ii For the LU23 LU28 LU29 series use a reducer bushing such as LM52 2400 3 T x 2 T fitting or the LM52 2410 3 S x 2 T fitting d The pipe length should run the measurement span and the bottom of the pipe should remain submerged at all times to prevent foam from entering the pipe e Cut a 45 notch at the bottom of the pipe and drill a 1 4 pressure equalization hole in the dead band f The pumps shoul...

Страница 19: ...he filtering averaging of the analog output Use this filter if the 4 to 20 mA output requires a smooth output for the application such as open channel flow measurement Decrease Output Filtering Placing a check mark in the box will eliminate all filtering averaging of the analog output Enables a pulse by pulse level reading Use this filter to see changes in level after every sound pulse Note Never ...

Страница 20: ...s tab has two key features The first is the ability to update the WebCal software to its latest version The second is the ability to update EchoSonic II s firmware to its latest version Demo The Demo tab has the ability to test the EchoSonic II sensor against a fixed target and can be used to confirm the switching points for the relays Note this feature is only a simulation and the relays are not ...

Страница 21: ...ory setting without configuration with WebCal Use the table below to identify the configuration of the sensor out of the box Factory Defaults Table Part Number Sensor Height Fill Height Loop Fail Safe Output at Empty Startup Condition LU27 _ _ Series 300 cm 118 1 290 cm 114 1 Overfill 22 mA 4 mA Bottom Empty LU23 _ _ Series 550 cm 216 5 530 cm 208 7 Overfill 22 mA 4 mA Bottom Empty LU28 _ _ Series...

Страница 22: ...allation level or application event for more specific troubleshooting direction User Settings Fill out the chart below and keep as a record of your configuration Tank Height Fill H Loop Fail Safe select one Hold Last Value Empty Full Overfill 21mA Overfill 22mA Output at Empty select one 4mA at Bottom 20mA at Bottom Startup Condition select one Empty Mid Tank 12mA Full Tank Overfill 22mA Advanced ...

Страница 23: ... corresponds to the full level of liquid from the bottom up and not the distance from the transmitter to the liquid top down Output of transmitter is opposite of the level of liquid Check the Output at Empty Setting No Unit Detected in WebCal WebCal cannot detect an EchoSonic II connected to the computer Check that the Fob is connected to the USB port Check that all four wires Red Black White and ...

Страница 24: ...procedures outlined above 2 have been subjected to electrical mechanical or chemical damage due to improper accidental or negligent use 3 have been modified or altered 4 anyone other than service personnel authorized by Flowline have attempted to repair 5 have been involved in accidents or natural disasters or 6 are damaged during return shipment to Flowline Flowline reserves the right to unilater...

Отзывы: