APPENDIX (continued)
Step Nine
Testing the Transmitter
1.
Connect a multimeter in series with the black wire to read the current output.
2.
Verify that the current increases (tank filling) and decreases (tank emptying) appropriately in the
calibrated span.
3.
If not, carefully observe and attempt to correlate any installation, level or application event for more
specific troubleshooting direction.
User Settings:
Fill out the chart below and keep as a record of your configuration.
Tank
Height =
Fill‐H =
Loop Fail‐Safe (select one)
Hold Last Value
Empty
Full
Overfill (21mA)
Overfill (22mA)
Output at Empty (select one)
4mA at Bottom
20mA at Bottom
Startup Condition (select one)
Empty
Mid‐Tank (12mA)
Full Tank
Overfill (22mA)
Advanced (if selected, identify which filter that was activated)
Increase output filtering:
Decrease output filtering:
Stabilize output in dead band