50
MN310140 Rev A2
Appendix
Section Nine
FACTORY SETTINGS
EchoPod
®
Sensor Height
Fill-Height
UG06 Series
236.2” (600 cm)
228.2” (579.7 cm)
UG12 Series
472.4” (1200 cm)
454.4” (1154.3 cm)
USER SETTINGS
Fill out the chart below and keep as a record of your configuration.
Tank
Height =
Fill-H =
Units
Inches
Feet
cm
Meter
Display
Air
Liquid
Safe
22mA
21 mA
20mA
Hold Last
4mA
Rev mA
4mA @ Bottom
4mA at Top
Start-up
4mA
12mA
20mA
22mA
TROUBLESHOOTING
PROBLEM
SOLUTION
No Unit Detected in
WebCal
®
.
If WebCal
®
cannot detect EchoPod
®
when connected to the computer:
Check that the LI99-2001 Fob is connected to the USB
®
port.
Check that all four wires (Red, Black, White and Green) are securely
attached to the Fob.
Check Device Manager and confirm that both drivers (WebCal
®
Configuration & EchoFob) are present.
If the above do not work, reinstall the WebCal
®
software while the computer
is in Safe Mode.
Cannot access some of
the configuration features
in WebCal
®
.
As choices are made in Configuration, WebCal
®
will begin to eliminate
functions that are not applicable to a configuration. To reset Configuration or
get access to all the features, click on the
Clear Screen
button.
Internet error. The server
name or address could not
be resolved.
This is a warning indicating that the computer configuring EchoPod
®
is not
connected to the Internet. Click
OK
to continue. Flowline recommends being
connecting to the Internet during configuration. But, not being connected to the
Internet will not prevent EchoPod
®
from being configured.