Table 1 Display Resolutions
Sensor pressure range (inches H
2
O)
Unit
±0.25
±0.5
±1.0
±5
±12
±20
Pa
0.01
0.1
0.1
1
1
1
kPa
0.00001
0.0001
0.0001
0.001
0.001
0.001
inH
2
O
0.0001
0.0001
0.001
0.001
0.01
0.01
mmHg
0.0001
0.0001
0.001
0.001
0.01
0.01
inHg
0.00001
0.00001
0.00001
0.0001
0.0001
0.001
psi
0.000001
0.00001
0.00001
0.0001
0.0001
0.0001
mbar
0.0001
0.001
0.001
0.01
0.01
0.01
lb/ft
2
0.001
0.001
0.001
0.01
0.01
0.1
m/sec
0.01
0.01
0.01
0.01
0.1
0.1
ft/sec
0.01
0.01
0.1
0.1
0.1
0.1
ft/min
1
1
1
1
1
1
m
3
/sec
0.0001
0.0001
0.0001
0.0001
0.0001
0.0001
ft
3
/sec
0.0001
0.0001
0.0001
0.0001
0.0001
0.0001
ft
3
/min
0.001
0.001
0.001
0.001
0.001
0.001
Table 2 Maximum and Minimum Transducer Speed Ranges (assuming
standard conditions, Patm = 29.92inHg and T = 70ºF) Using Pitot Static Probe
Pressure Transducer Range
Speed Range
±0.25 inH
2
O (±60 Pa)
100 - 2000 ft/min (0.5- 10.2 m/sec)
±0.5 inH
2
O (±124 Pa)
141 - 2829 ft/min (0.7- 14.4 m/sec)
±1 inH
2
O (±249 Pa)
200 - 4001 ft/min (1 - 20.3 m/sec)
±5 inH
2
O (±1.2 kPa)
447 - 8946 ft/min (2.3 - 45.4 m/sec)
±12 inH
2
O (±3 kPa)
693 - 13859 ft/min (3.5 - 70.4 m/sec)
±20 inH
2
O (±4.98 kPa)
895 - 17892 ft/min (4.5 – 90.9 m/sec)