If the measured pressures fluctuate more than 10%, an indicator
T
will appear at
the top RHS of the display. This may suggest that another measurement location
may be necessary.
Q: What Probes can I use with the FKS 1DP-PBM-E?
Any Pitot static probe or S type probe can be used. Most Pitot probes have unity
calibration coefficients (K factor). S type probes (used for
dirty
air streams) have
coefficients around 0.80 to 0.92. To set K, see below.
Q: How Do I Connect the Pitot Probe?
See the sketch below. P+ is the top left barb; P- is the top right barb. For
connection of other types of probes, the stagnation (total) pressure port should be
connected to the P+ barb; the static or reference pressure should be connected to
the P- barb.
Q: How Do I Change K?
To set K, or a probe’s calibration constant (default value = 1), set the Range
selector to Menu. The menu will cycle through various functions. When
Set K
factor
appears, rotate (any direction) the Range control to select it. The
menu will prompt,
Set K Rotate
followed by
K factor
. Rotate the
Range selector (any direction or unit).
Press & Rotate
will be briefly
displayed. The screen will display
0.00
. The first digit will begin to blink. To
increment this digit press the Hold/Zero button. You can also keep the button
pressed to advance rapidly. If the displayed digit increments beyond 9, it will wrap
around and continue from 0. When the desired value for this digit is set, rotate the
Range selector (any direction or unit). The second digit will begin to blink. Use the