Measuring Flow Rate
Flow rate is measured by calculating an average speed for the conduit of interest,
and then, multiplying this speed by the cross sectional area of the duct at the
measurement location. The speed value may be estimated using a single reading
or a survey across the duct at a station (see Appendix A).
The Menu System
The menu system uses prompt driven operation for ease of use. All option
selections are performed by rotating the Range selector (in any direction and to
any unit). This provides positive tactile and audible feedback that your selection
has been made. All numbers are set by pressing the Hold/Zero button to
increment the digit and by rotating the Range selector to accept the displayed
number and advance to the next digit. Options available through using the menu
system are:
Set Patm & T
o
Used to set actual speed
Set K factor
o
Used to set the calibration constant (K) for non-unity
probes (e.g. S type probes)
View K, Patm & T
o
Used to view the currently stored values for these
parameters
Reset K, Patm, T
o
Used to reset to default and standard conditions (K=1,
Patm=29.92inHg and T=70ºF)
Smooth/Average
o
Used to select between exponential smoothing or
averaging for measurements