4.8 Communication Setting Interface
Click
System
Settings
button in the main interface, then click
Communication
button
to enter Communication Settings interface.
This pump supports MODBUS-RTU Mode. Please select baud rates and
communication interface (RS485/RS232). Click
Slave No.
button to enter peristaltic
pump address No. (range:1-32). Turn the communication enable toggle to
ON
. This
pump can now communicate with master, receiving master signal.
Note
: After settings, the peristaltic pump only receives communication control when
in the main interface. It is invalid for the communication control in the other settings
interface.