EM6100 Handheld Portable Ultrasonic Flowmeter Ver.8.08
Flomotion
Systems,
Inc.
EM6100
41
for the purpose of the compatibility with the former versions
(4)
Command Connector &
The & command connector can connect up to 6 basic commands to form a longer command so that it
will make the programming much easier.
For example, assume that the measurement of an instrument with IDN=4321 are going to be returned,
and (then) all the following 3 values--- (1) flow rate (2) velocity (3)POS totalizer---will be returned
simultaneously. The combined command would be W4321DQD&DV&DI+(CR), and the result would
be:
+1.12m3/d(CR)
+3.100m/s(CR)
+10m3(CR)
§6.4 Codes for the Keypad
The codes for the keypad should be used when the instrument is connected with other terminals that
operate the instrument by transmitting the ‘M’ command along with the keypad code. By this function,
remote operation of this instrument can be realized, even via the Internet.
Key
Hexadecimal
Key code
Decimal
Key code
ASCII
Code
Key
Hexadecim
al Key code
Decimal
Key code
ASCII
Code
0 30H 48 0
8 38H 56 8
1 31H 49 1
9 39H 57 9
2 32H 50 2
. 3AH 58 :
3 33H 51 3
◄
3BH,0BH 59
;
4 34H 52 4
MENU
3CH,0CH
60 <
5 35H 53 5
ENT
3DH,0DH
61 =
6 36H 54 6
▲
/+ 3EH 62 >
7 37H 55 7
▼
/- 3FH 63 ?