EM6100 Handheld Portable Ultrasonic Flowmeter Ver.8.08
Flomotion
Systems,
Inc.
EM6100
32
So there are 36 different flow rate units in total for selection.
M32
Window for selecting the totalizier working unit
M33
Select totalizer multiplier
The multiplier ranges from 0.001 to 10000
M34
Turn on or turn off the NET totalizer
M35
Turn on or turn off the POS totalizer
M36
Turn on or turn off the NEG totalizer
M37
(1) Totalizer reset
(2) Restore the instrument to the default parameters as the manufacturer did by
pressing the dot key followed by the backspace key.
Take care or make note of the
parameters before doing the restoration
M38
Press-a-key-to-run or to stop totalizer for easier calibration
M39
Operational interface language selection in English or Chinese.
M40
Flow rate damping for a stable value. The input range is 0 to 999 seconds.
0 means there is no damping. Default value is 10 seconds
M41
Lower flow rate cut-off to avoid invalid accumulation.
M42
Zero point setup under the condition when there is no liquid running inside the pipe.
M43
Clear the zero point set by the user, and restore the zero point set by the
manufacturer
M44
Set up a manual flow bias. Generally this value should be 0.
M45
Scale factor for the instrument. The default value is ‘1’.
Keep this value as ‘1’, when no user calibration has been made.
M46
Network environment Identification Number. Any integer can be entered except
13(0DH, carriage return), 10 (0AH, line feeding), 42 (2AH), 38, 65535.
Every set of the instrument in a network environment should have a unique IDN.
Please refer to the chapter for communication.
M47
System locker to avoid modification of the parameters
M48
Not used
M49
Communication tester
M50
“Option” selection for the built-in logger. It also functions as the switch of logger
M51
Time setup for the data logger
M52
(1) Data logging direction control. If ‘To RS-232’ is selected, all the data produced
by the data logger will be transmitted out through the RS-232 interface
(2) If ‘To buffer ‘ is selected, the data will be stored into the built-in logger memory
(3) Buffer transferring and buffer clearing
M53
Logger buffer viewer. It functions as a file editor. Use Dot, backspace UP and DN
keys to browse the buffer.
If the logger is ON, the viewer will automatically refresh once new data are stored.
M61
Display Version information and Electronic Serial Number (ESN) that are unique