Fixed frequency mode is used
for service purposes. Required
frequency is set directly in Hz in
range 1-120000 Hz in position 4
and 1-1200 Hz in position 5.
·
Negativ e/n o n - n eg at i v e
flow rate
This mode is used for flow di-
rection differentiation. Output is
closed/open for negative flow
rate.
·
Failure occurred/not oc-
curred
If failure with mode set to active
(see
error messages
) occurs,
the output closes/opens for at
least 5 seconds. If failure per-
sists, output is closed/open for
whole period of failure duration.
·
Limit flow rate exceeded/
undergone (not)
If flow rate is higher/lower than
limit value set, the output closes
(opens). When flow rate returns
to limits, the output opens
(closes) again taking in account
preset hysteresis. This function
Fig. Q > Qlimit
Fig. Q < Qlimit
Fig. IQI > Qlimit
Fig. IQI < Qlimit
Fig. Menu structure of B modules
Fig. Terminal connection
6 7
B1
B
in
ar
y
ou
tp
ut
O
pt
oM
O
S
3
50
V
8 9
B2
B
in
ar
y
ou
tp
ut
ac
tiv
e
5
V/
1
00
m
A
+
-
8 9
B5
B
in
ar
y
ou
tp
ut
R
e
la
y
4.Binary output
─► ●
Permanently open
○
Perm. Closed
○
IQI pulses
» Pulse length [10ms]
○
IQI not pulses
» Pause length [10ms]
○
Q+ pulses
» Vol./ pulse [ml]
○
Q+ not pulses
○
Q- pulses
○
Q- not pulses
○
Q+ frequency
» Flow rate at 1 kHz
○
Q- frequency
○
IQI frequency
○
Fixed frequency
» Fixed frequency [ Hz ]
○
Negative flow
○
Non-neg. flow
○
'Error occured
√
E1
○
'No error occured
√
…
√
E13
○
Q>Qlim.
» Flow limit
○
Q>Qlim. not
» Hysteresis
○
Q<Qlim.
○
Q<Qlim. not
○
IQI>Qlim.
○
IQI>Qlim. not
○
IQI<Qlim.
○
IQI<Qlim. not
○
Cleaning
○
Not cleanig
○
Batch Opened
» Volume Advance [ml]
○
Batch Opened not
» Time Advance [ms]
○
Batch Stop puls
» Pulse length [10ms]
○
Batch Stop /puls
» Volume Advance [ml]
» Time Advance [ms]
works in four modes of depend-
ence on flow rate with output
polarity differentiation.
·
Electrode
cleaning/no
cleaning
The output is closed/open dur-
ing cleaning.
·
Batch Opened
·
Batch /Opened
Output is switched on /
switched off during the batch
running. It is possible to set the
advance of the output before
batch ending. Advance could be
set by time or volume. Detailed
information is given in chapter
Batching.
·
Batch Stop pulse
·
Batch Stop /pulse
Output generates pulse for
ending of the batch. It is possi-
ble to set pulse length (10ms -
2,5s) and pulse advance before
batch ending. Advance could be
set by time or volume. Detailed
information is given in chapter
Batching.