8.10
IRIG-B
The FLIR X6520sc features an IRIG-B input BNC connector (colored blue) on the back
panel of the camera (8 in Figure 5.3
Camera back panel description.
The value of the signal is digitalized during the integration of the infrared image and em-
bedded within it. It is then recorded by ResearchIR Max and stored in the sequence file
image headers.
The supported IRIG-B formats are IRIG-B12x. The signal should be 3:1, 3 Vpp maximum
at 50 Ω or 6 Vpp for high impedance input.