8.6.2
Camera setup when using the Protective Window (P/N 1 194 977)
The protective window (P/N 1 194 977) contains an optical material that affects the
transmission of infrared radiation to the FPA detector inside the camera. This means
that you have to specify a temperature and a transmission value for external optics
in the camera software for P and S series cameras.
Follow this procedure to enter the temperature and transmission value for external
optics:
Action
Step
Point to
Analysis
on the menu bar and press the joystick.
1
Point to
Object param
and press the joystick.
2
Set
External optics
to
On
.
3
Enter a transmission value of 0.83 in the
Optics transmission
text box by moving
the joystick left/right. This value has been measured at FLIR Systems AB, Sweden.
4
Enter an external temperature for the lens in the
Optics temperature
text box by
moving the joystick left/right. Usually, this temperature is the same temperature
as the camera’s ambient temperature. However, in some situations – such as when
looking at very hot targets – the temperature can be considerably higher.
5
Press the joystick to confirm the changes and leave the dialog box.
6
8.6.3
Focusing the camera using autofocus
Action
Step
Press the green ON/OFF button to switch on the camera.
1
Press and hold down the
A
button for one second to adjust the focus. An indicator
will be displayed on the left side of the screen when focusing.
2
8.6.4
Focusing the camera manually
Action
Step
Press the green ON/OFF button to switch on the camera.
1
Adjust the focus by moving the joystick up/down. An indicator will be displayed
on the left side of the screen when focusing.
2
8.6.5
Using the electronic zoom
Action
Step
Press the green ON/OFF button to switch on the camera.
1
8
Publ. No. 1557978 Rev. a155 – ENGLISH (EN) – February 6, 2006
49
8 – Tutorials
Содержание ThermaCam P25
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ...ThermaCAM P25 User s manual Publ No 1557978 Rev a155 ENGLISH EN February 6 2006...
Страница 10: ...Reg No Status Designation 29 233 400 Pending U S x Publ No 1557978 Rev a155 ENGLISH EN February 6 2006...
Страница 16: ...INTENTIONALLY LEFT BLANK 1 2 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 1 Warnings cautions...
Страница 26: ...INTENTIONALLY LEFT BLANK 4 12 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 4 Packing list...
Страница 66: ...INTENTIONALLY LEFT BLANK 8 52 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 8 Tutorials...
Страница 78: ...INTENTIONALLY LEFT BLANK 9 64 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 9 Camera overview...
Страница 98: ...INTENTIONALLY LEFT BLANK 11 84 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 11 Folder and file structure...
Страница 104: ...INTENTIONALLY LEFT BLANK 12 90 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 12 Electrical power system...
Страница 108: ...INTENTIONALLY LEFT BLANK 14 94 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 14 Maintenance cleaning...
Страница 158: ...INTENTIONALLY LEFT BLANK 20 144 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 20 Theory of thermography...
Страница 191: ......