Thermographic measurement techniques
23
5. Measure the apparent temperature of the aluminum foil and write it down.
Figure 23.5
Measuring the apparent temperature of the aluminum foil.
23.2.1.2
Step 2: Determining the emissivity
Follow this procedure:
1. Select a place to put the sample.
2. Determine and set reflected apparent temperature according to the previous
procedure.
3. Put a piece of electrical tape with known high emissivity on the sample.
4. Heat the sample at least 20 K above room temperature. Heating must be reasonably
even.
5. Focus and auto-adjust the camera, and freeze the image.
6. Adjust
Level
and
Span
for best image brightness and contrast.
7. Set emissivity to that of the tape (usually 0.97).
8. Measure the temperature of the tape using one of the following measurement
functions:
•
Isotherm
(helps you to determine both the temperature and how evenly you have
heated the sample)
•
Spot
(simpler)
•
Box Avg
(good for surfaces with varying emissivity).
9. Write down the temperature.
10. Move your measurement function to the sample surface.
11. Change the emissivity setting until you read the same temperature as your previous
measurement.
12. Write down the emissivity.
Note
• Avoid forced convection
• Look for a thermally stable surrounding that will not generate spot reflections
• Use high quality tape that you know is not transparent, and has a high emissivity you
are certain of
• This method assumes that the temperature of your tape and the sample surface are
the same. If they are not, your emissivity measurement will be wrong.
23.3
Reflected apparent temperature
This parameter is used to compensate for the radiation reflected in the object. If the
emissivity is low and the object temperature relatively far from that of the reflected it will
be important to set and compensate for the reflected apparent temperature correctly.
#T559950; r. AD/35720/35720; en-US
89
Содержание A615
Страница 1: ...User s manual FLIR A6xx series...
Страница 2: ......
Страница 3: ...User s manual FLIR A6xx series T559950 r AD 35720 35720 en US iii...
Страница 4: ......
Страница 8: ......
Страница 22: ...Mechanical installation 9 Example 2 cable strain relief with cable clamps T559950 r AD 35720 35720 en US 14...
Страница 71: ...Mechanical drawings 16 T559950 r AD 35720 35720 en US 63...
Страница 81: ...CE Declaration of conformity 17 T559950 r AD 35720 35720 en US 73...
Страница 82: ......
Страница 84: ...Digital I O connection diagrams 19 T559950 r AD 35720 35720 en US 76...
Страница 85: ......