Technical data
14
14.8
FLIR A615 windowing 80°
P/N: 55001-0105
Rev.: 35207
General description
The FLIR A615 has features and functions that make it the natural choice for anyone who uses PC soft-
ware to solve problems and needs 640 × 480 pixel resolution. Among its main features are GigE Vision
and GenICam compliance, which makes it plug-and-play when used with software packages such as
IMAQ Vision and Halcon.
The camera is equipped with a 80° lens.
Key features:
•
Affordable.
•
GigE compliant.
•
GenICam compliant.
•
Trigg/synchronization/GPIO.
•
16-bit 640 × 480 pixel images at 50 Hz, signal, temperature linear, and radiometric.
•
Windowing mode: 640 × 240 pixels at 100 Hz or 640 × 120 pixels at 200 Hz.
•
Compliant with any software that supports GenICam, including National Instruments IMAQ Vision
and Stemmers Common Vision Blox.
•
Open and well-described TCP/IP protocol for control and set-up.
Typical applications:
•
High-end infrared machine vision that requires temperature measurement.
•
Slag detection.
•
Food processing.
•
Electronics testing.
•
Power resistor testing.
•
Automotive.
Imaging and optical data
IR resolution
640 × 480 pixels
Thermal sensitivity/NETD
< 0.05°C @ +30°C (+86°F) / 50 mK
Field of view (FOV)
80° × 64.4° (92.8° diagonal)
Minimum focus distance
65 mm (2.6 in.)
Focal length
6.5 mm (0.26 in.)
Spatial resolution (IFOV)
2.62 mrad
Lens identification
Automatic
F-number
1.0
Image frequency
50 Hz (100/200 Hz with windowing)
Focus
Automatic or manual (built in motor)
Detector data
Detector type
Focal plane array (FPA), uncooled
microbolometer
Spectral range
7.5–14 µm
Detector pitch
17 µm
Detector time constant
Typical 8 ms
Measurement
Object temperature range
•
–40°C to +150°C (–40°F to +302°F)
•
100 to +650°C (+212 to +1202°F)
•
300 to +2000°C (+572 to +3632°F)
Accuracy
±2°C (±3.6°F) or ±2% of reading
#T559950; r. AD/35720/35720; en-US
38
Содержание A615
Страница 1: ...User s manual FLIR A6xx series...
Страница 2: ......
Страница 3: ...User s manual FLIR A6xx series T559950 r AD 35720 35720 en US iii...
Страница 4: ......
Страница 8: ......
Страница 22: ...Mechanical installation 9 Example 2 cable strain relief with cable clamps T559950 r AD 35720 35720 en US 14...
Страница 71: ...Mechanical drawings 16 T559950 r AD 35720 35720 en US 63...
Страница 81: ...CE Declaration of conformity 17 T559950 r AD 35720 35720 en US 73...
Страница 82: ......
Страница 84: ...Digital I O connection diagrams 19 T559950 r AD 35720 35720 en US 76...
Страница 85: ......