Technical data
14
Measurement analysis
Atmospheric transmission correction
Automatic, based on inputs for distance, atmos-
pheric temperature and relative humidity
Optics transmission correction
Automatic, based on signals from internal sensors
Emissivity correction
Variable from 0.01 to 1.0
Reflected apparent temperature correction
Automatic, based on input of reflected
temperature
External optics/windows correction
Automatic, based on input of optics/window trans-
mission and temperature
Measurement corrections
Global object parameters
USB
USB
•
Control and image
USB, standard
USB 2 HS
USB, connector type
•
USB Mini-B
USB, communication
TCP/IP socket-based FLIR proprietary
USB, image streaming
16-bit 640 × 480 pixels @ 25 Hz
•
Signal linear
•
Temperature linear
•
Radiometric
USB, protocols
TCP, UDP, SNTP, RTSP, RTP, HTTP, ICMP, IGMP,
ftp, SMTP, SMB (CIFS), DHCP, MDNS (Bonjour),
uPnP
Ethernet
Ethernet
Control and image
Ethernet, type
Gigabit Ethernet
Ethernet, standard
IEEE 802.3
Ethernet, connector type
RJ-45
Ethernet, communication
TCP/IP socket-based FLIR proprietary and GenI-
Cam protocol
Ethernet, image streaming
16-bit 640 × 480 pixels @ 50 Hz
16-bit 640 × 240 pixels @ 100 Hz
16-bit 640 × 120 pixels @ 200 Hz
•
Signal linear
•
Temperature linear
•
Radiometric
GigE Vision and GenICam compatible
Ethernet, protocols
TCP, UDP, SNTP, RTSP, RTP, HTTP, ICMP, IGMP,
ftp, SMTP, SMB (CIFS), DHCP, MDNS (Bonjour),
uPnP
Digital input/output
Digital input, purpose
Image tag (start, stop, general), Image flow con-
trol, (stream on/off), Input ext. device (program-
matically read)
Digital input
2 opto-isolated, 0–1.5 V = low, 3–25 V = high
Digital output, purpose
Output to ext. device (programmatically set)
Digital output
2 opto-isolated, ON = supply (max. 100 mA), OFF
= open
Digital I/O, isolation voltage
500 VRMS
#T559950; r. AD/35720/35720; en-US
27
Содержание A615
Страница 1: ...User s manual FLIR A6xx series...
Страница 2: ......
Страница 3: ...User s manual FLIR A6xx series T559950 r AD 35720 35720 en US iii...
Страница 4: ......
Страница 8: ......
Страница 22: ...Mechanical installation 9 Example 2 cable strain relief with cable clamps T559950 r AD 35720 35720 en US 14...
Страница 71: ...Mechanical drawings 16 T559950 r AD 35720 35720 en US 63...
Страница 81: ...CE Declaration of conformity 17 T559950 r AD 35720 35720 en US 73...
Страница 82: ......
Страница 84: ...Digital I O connection diagrams 19 T559950 r AD 35720 35720 en US 76...
Страница 85: ......