FAA-27, Technical Manual, Rev. 1.1, June 2014
Page 15 of 21
2. eCal procedure steps:
Descriptions
Step 1 START eCal
Command:
@011000000001020001EB
Start the eCal process.
Response:
@011000000001EE
ACK
Step 2. Total Load cell Emax entry
Command:
@01100003000204000001F4F1
Total LC is 500kg.
Response:
@011000030002EA
ACK
Step 3. Average Load cell mV/V entry
Command:
@0110000500020400004E2F67
LC sensitivity: 2.0015
mV/V
.
Response:
@011000050002E8
ACK
Step 4. Estimated Dead load entry
Command:
@0110000700020400000000E2
Dead load is 0kg.
Response:
@011000070002E6
ACK
Step 5. Scale Capacity entry
Command:
@01100009000204000000FAE6
Scale capacity is 250kg.
Response:
@011000090002E4
ACK
Step 6. SAVE with estimated dead load value
Command:
@011000000001020007E5
Save the eCal values.
Response:
@011000000001EE
ACK
Or SAVE with automatic zero adjustment.
Unload the scale and send the command below
to start the zero adjustment.
Command:
@011000000001020015D7
Save eCal with Zero Adj
Response:
@011000000001EE
ACK
Step 7. APPLY
Command:
@011000000001020009E3
Apply the calibration.
Response:
@011000000001EE
ACK
Step 8. STOP eCal
Command:
@011000000001020002EA
Stop the eCal process
Response:
@011000000001EE
ACK
3. Digital Filter Setting:
Command:
@011000220001020000CA
Set the Fast filter
Command:
@011000220001020001C9
Set to Medium filter
Command:
@011000220001020002C8
Set to Slow filter
Response:
@011000220001CC
ACK