C.4.12 Commands Analogue Output 4 – 20 mA - AL, AH, AA
The following commands must be saved in EEPROM by command AS.
AA
Sets the source which will be used by analogue output
Possible selections for analogue output:
AA = 0
gross weight
AA = 1
net weight
AA = 2
Peak value
AA = 3
Average value
AA = 4
Hold value
AA = 5
Peak-to Peak value
AA = 6
Valley value
AA = 7
Hold value
AA = 8
analogue output OFF
Request / Set up analogue output:
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
AA
¿
A+00001
set net
AA_0
¿
OK
set gross
Factory default: 2 [= analogue output OFF]
AH
Set the analogue high level
Request / Set up high level for analogue output 20 mA:
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
AH
¿
H+10000
Request: 10 000 d @ 20 mA
AH_5000
¿
OK
Setting: 5 000 d @ 20 mA
Factory default: 10 000 [10 000 d = 20 mA]
AL
Set the analogue low level
Request / Set up low level for analogue output 4 mA:
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
AL
¿
L+00000
Request: 00 000 d @ 4 mA
AL_00500
¿
OK
Setting: 500 d @ 4 mA
Factory default: 00 000 [0 d = 4 mA]