background image

 

1

Flinn Goggle Sanitizer

 Instruction Manual

Introduction

Congratulations on purchasing the finest goggle sanitizer made. We know you'll be pleased with our goggle sanitizer's crafts-

manship and durability!

Receiving

Please unpack the goggle sanitizer slowly and carefully. Once unpacked, verify that all the component parts are accounted for: 

  •     1 each   Flinn Goggle Sanitizer
  •     1 each   25-watt germicidal lamp (installed in sanitizer)
  •     4 each   Rubber feet for countertop use
  •     2 each   Door lock keys

Please advise us of any breakage or shortage by calling 1-800-452-1261. No sale is complete unless you are satisfied!

Instructions

1.  The germicidal lamp is secured in place with temperature rated electrical tape for shipping purposes only. The tape is not 

necssary for operation. It can be removed or left in place. Please verify that the germicidal lamp did not dismount during 

shipping. The lamp should be secured in each socked with the pins seated horizontally.

2.  If the goggle sanitizer is to be placed on the top of a bench or table, the four rubber feet should be used. Apply a rubber 

foot to each corner of the bottom of the cabinet.

3.  The goggle sanitizer may be all mounted using the two keyhole-shaped mounting holes in the upper back of the cabinet 

(covered by white tape). Two additional screws will be needed to secure the lower portion of the cabinet against the wall 

and to prevent the cabinet from being lifted off the upper set of screw. Pretest the goggle sanitizer prior to wall mounting.

4.  Plug the goggle sanitizer into a properly grounded receptacle only.
5.  Place each pair of safety goggles of glasses in a separate bin, centered front to back. The goggles should be placed in the 

bin with the lens down. This will prevent any slippage of the goggles. Load the goggles from top to bottom so that less 

than full loads can benefit from shorter UV exposure times. The surfaces of the eyewear that touch the skin should face 

the light source (i.e., place the eyewear with the lenses down). 

6.  Close and lock the doors. Door interlock safety switches will not allow the unit to operate unless both doors are fully 

closed. Do not operate this unit if these switches are not functioning properly. Contact Flinn Scientific for switch replace-

ment.

7.  The goggle sanitizer can be started by: 

  

•  Analog Model: Turning the timer know clockwise to the desired exposure time.

   

•  Digital Model: Depressing the power switch and setting the desired exposure time.

8.  If time is available between each use of safety goggles, a UV dose of 15 minutes is recommended for optimum results, 

especially at full capacity. Five minutes is the minimum effective dosage time. Look through the viewing lens on the front 

of the cabinet to ensure that the UV lamp is functioning with each use.

9.  This goggle sanitizer is equipped with a 25-watt germicidal lamp that is rated at 5,000 hours of continuous use. Short 

periods of on/off cycling, as in this application, will reduce lamp life. With normal use of this lamp, the ultraviolet por-

tion of the lamp will wear out before the visible portion, further reducing the effective lamp life. In order to maintain 

optimum performance, it is recommended that the lamp be replaced after 2,000 to 3,000 hours of use. As an example, if 

the unit is run for 15 minutes, 5 times a day, 5 days a week, for 40 weeks, the lamp should be replaced in approximately 8 

years. A lamp replacement record has been provided on the inside surface of the right-hand door. Write in the date of the 

first use and subsequent lamp changes. If a new lamp is needed, please order Catalog No. SE1001.

10.  Please remember that exposure to UV light can cause vision loss. Prolonged exposure to ultraviolet light causes con-

junctivitis, cataracts and/or permanent damage to the retina. Do not attempt to operate the unit with the doors open or 

unlocked.

© 2018 Flinn Scientific, Inc. All Rights Reserved.

Publication No. 12835 

IN12835 / 051718