![Flexim PIOX R500 Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/flexim/piox-r500/piox-r500_operating-instruction_2304740081.webp)
15 Data logger
PIOX R721
15.1 Configuration of the data logger
81
UMPIOX_F72xV1-3EN, 2021-09-01
Storage rate
If a starting time for storing the measured values is defined, a storage rate has to be entered. The storage rate indicates
the interval how often the measured values are transmitted or stored.
• Select the menu item
Storage rate
.
• Press ENTER.
• Select a storage rate from the scroll list.
• Press ENTER.
• If
Customized
is selected, enter the storage rate.
• Press ENTER.
Ringbuffer
The data logger can be configured as linear logger or ringbuffer. If the ringbuffer is deactivated and the data logger is full,
the storing of measured values will be terminated. It can be continued after clearing the data logger. If the ringbuffer is
activated and the data logger is full, the oldest measured values will be overwritten. In ringbuffer mode, the remaining
capacity of the data logger is displayed during the measurement, e.g.:
Log←→ : 1d 6h 57m
is displayed, if no measured values have been overwritten.
Log|←→| : 1d 6h 57m
is displayed, if the old measured values have been overwritten.
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Ringbuffer
is displayed.
• Select
On
to activate the ringbuffer.
• Press ENTER.
If the ringbuffer is deactivated and the data logger is full, the storing of measured values will be terminated.
• Select
Off
to deactivate the ringbuffer.
• Press ENTER.
Storage mode
• Select the menu item
Storage mode
.
• Press ENTER.
• Select
Sample
to store the current measured value.
• Select
Average
if the average of all undamped measured values of a storage interval is to be stored.
Data logger\Configuration\Storage rate
Data logger\Configuration\Ringbuffer
Data logger\Configuration\Storage mode
Notice!
The storage mode does not affect the outputs.
Notice!
Storage mode
=
Average
The average of the physical quantity and of other quantities (e.g., the measured temperature) will be calculated.
If a storage rate < 5 s is selected,
Sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as invalid.