7 Connection
66
UMFLUXUS_F7V4-6-2EN, 2017-10-01
7.1.3.3
Transducer Cable with Stainless Steel Conduit and Stripped Cable Ends
• Remove the right blind plug for the connection of the transducer cable (see Fig. 7.22).
• Insert the transducer cable into the junction box.
• Fix the transducer cable by tightening the cable gland.
• Connect the transducer cable to the terminals of the junction box (see Fig. 7.21, Fig. 7.22 and Tab. 7.10).
Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Tab. 7.10: Terminal assignment (KL1)
terminal (JB01)
connection
V
transducer (white cable, marked white)
VS
transducer (red cable)
RS
transducer (red cable)
R
transducer
(white
cable)
Fig. 7.22: Connection of the transducer cable
with stainless steel conduit and stripped cable ends
V VS
R
RS
cable gland
junction box JB01