Flashback Technologies CipherOx CRI M1 Скачать руководство пользователя страница 24

Page 

24

 of 

24

 

Figure 14 

 Box and Whisker plots of one-minute average CRI values, before blood loss, at maximum blood 

loss  (at  end  of  blood  withdrawal  for  subjects  completing  ~20%  blood  volume  withdrawal),  during 
symptoms (at  end of blood withdrawal for subjects experiencing symptoms and not  completing ~20% 
blood loss), and after blood replacement. For each panel above, the number of subjects was 42, 32, 7, and 
32, respectively. 
 

Conclusions 

The results of this study met the pre-specified pass/fail criteria and thus provide supporting evidence that 
CRI  trends  with  intravascular  volume  changes  as  compared  to  direct  measurement  of  blood  volume 
decrease and increase. It should be noted that this clinical validation of CRI was performed using healthy 
volunteers aged 19-36 years under supine, non-motion conditions.  

Clinical Use Considerations 

Clinical users of the M1 should understand the basic principles regarding the derivation of the CRI value 
as  well  as  the  strengths  and  limitations  of  research  conducted  to  date.  As  noted  above,  in  normal 
volunteers under laboratory conditions, CRI was found to correlate closely with changes in intravascular 
blood volume (up to ~ 20% of blood volume loss). 

Nonetheless, this device is intended as an adjunct in 

patient assessment. It must be used in conjunction with other methods of assessing clinical signs and 
symptoms. 

 
CRI estimates are based on reference data collected in ideal conditions. Progressive declines to low CRI 
values  indicate  measurements  based  on  subjects  in  clinical  studies  exhibiting  symptoms  of  increasing 
hemodynamic instability, most  notably precipitous  hypotension and clinical symptoms associated with 
shock. However, since many known and unknown factors (artifacts, advanced age, cardiovascular disease, 
medications,  acute  trauma,  etc.)  may  affect  these  measurements,  changes  in  CRI  should  always  be 
interpreted cautiously as a 

potential

 indicator of changes in patient status due to volume change. It is 

recommended that users carefully consider the clinical conditions and evaluate conventional vital signs 
when interpreting changes in CRI.  

References  

Convertino, V, G. Grudic, J. Mulligan, and S. Moulton. 2013. "Estimation of individual-specific progression 

to impending cardiovascular instability using arterial waveforms." 

Journal of Applied Physiology

 

115 (8): 1196-202. 

 
Convertino, V, J Howard, C Laborde, S Cardin, P Batchelder, J Mulligan, G Grudic, and D Macleod. 2015. 

Individual-specific,  beat-to-beat  trending  of  significant  human  blood  loss:  The  compensatory 

reserve.”

 

SHOCK

 44: 27

32. 

 
Moulton, S, J Mulligan, M.A. Santoro, K Bui, G Grudic, and D MacLeod. 

2017. “

Validation of a noninvasive 

monitor to continuously trend indiv

idual responses to hypovolemia.”

 

Journal of Trauma and Acute 

Care Surgery

 83 (1): S104

S111. 

 
Moulton,  S.,  J.  Mulligan,  G.  Gr

udic,  and  V.  Convertino.  2013.  “

Running  on  empty?  The  compensatory 

reserve index.”

 

Journal of Trauma and Acute Care Surgery

 75 (6): 1053-1059. 

 

Содержание CipherOx CRI M1

Страница 1: ...shback M1 Compensatory Reserve Monitor Handheld with Flashback s Patented CRI Technology Instructions for Use Version 1 0 2 May 2020 Flashback Technologies Inc 1 Little West 12th Street Suite 505 New...

Страница 2: ...o Software Symbols and Parameters 10 Parameters 10 Graphical Interface Symbols 10 Gauge 11 Trend 11 Sensor Status 11 Screen Status 12 Battery Status 12 Adjusting the Screen Brightness 12 Powering off...

Страница 3: ...x CRI 19 Estimating CRI from Pulsatile Waveforms 20 Performance of CRI 20 Summary of Clinical Study 21 Background Information Regarding Use of CRI 21 Compensatory Reserve 21 Development of the Compens...

Страница 4: ...ients with a finger thickness of 0 3 to 1 in hospital and pre hospital settings CRI trends with changes in intravascular volume relative to the individual patient s response to hypovolemia and should...

Страница 5: ...d replace the sensor Do not use if any component of the device is visibly damaged Do not use if the battery door is broken or fails to close tightly as this may allow ingress or ESD damage to the devi...

Страница 6: ...ying it to a patient The Model M1 is non sterile Do not sterilize autoclave or immerse the Model M1 in liquid Do not pour or spray any liquids into the device Do not use caustic or abrasive cleaning a...

Страница 7: ...patient isolation from electrical shock UL Mark for Canada and the United States with respect to electric shock fire and mechanical hazards only in accordance with IEC 60601 1 AAMI ANSI 60601 1 and C...

Страница 8: ...angle up to 60 from the vertical and ingress of solid foreign objects greater than or equal to 2 5 mm in diameter per IEC 60529 Do not expose the device to a magnetic resonance MR environment Conventi...

Страница 9: ...he left side of the M1 with the Nonin logo facing up see Figure 2 Figure 2 Attaching the Nonin Sensor Using the Nonin Articulated Clip Sensor 1 Insert a finger preferably the index middle or ring fing...

Страница 10: ...ually available within 5 to 20 seconds of sensor placement Figure 4 Power on the M1 Guide to Software Symbols and Parameters Parameters HR Heart Rate SpO2 Functional oxygen saturation of arterial hemo...

Страница 11: ...ee Figure 6 If the sensor is not connected or there is no signal the trend will continue to move leaving empty space in the graph in order to indicate time has passed since data has been collected Fig...

Страница 12: ...battery Less than 20 minutes of device use may remain The battery status indicator is only valid when an alkaline battery is used Adjusting the Screen Brightness The screen starts up in Normal brightn...

Страница 13: ...displayed as _ Cleaning the Reusable Sensor Cautions Clean the sensor before applying it to a new patient Unplug the sensor from the M1 before cleaning Do not sterilize autoclave or immerse the senso...

Страница 14: ...household bleach containing less than 10 sodium hypochlorite Do not sterilize autoclave or immerse the M1 in liquid of any kind Do not pour or spray any liquids onto the M1 Do not sterilize with EtO D...

Страница 15: ...Page 15 of 24 Nonin OEM III Specifications The M1 utilizes the internal Nonin OEM III module to generate SpO2 measurements...

Страница 16: ...5 volt AA battery CRI is not displayed Ensure snug connection between M1 and sensor OR Ensure Sensor Status state bottom left corner of M1 display shows a completely green outline and does not displa...

Страница 17: ...Group 1 The M1 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11...

Страница 18: ...50 DC 27 450 FM 1 kHz Sine 5 Hz Deviation 28 710 745 780 Pulse 217 Hz 50 DC 9 810 870 930 Pulse 18 Hz 50 DC 28 1720 1845 1970 Pulse 18 Hz 50 DC 28 2450 28 5240 5500 5785 9 M1 Compliance The M1 is not...

Страница 19: ...l the patient is already near collapse and at risk for poor tissue perfusion progressive acidosis and sudden unexpected hemodynamic decompensation The Compensatory Reserve Index CRI provides a single...

Страница 20: ...case CRI is based on the IR PPG signal from the Nonin OEM III 8000AA Performance of CRI CRI is a new physiological monitoring parameter which trends with intravascular volume and can be used by clinic...

Страница 21: ...tudied and observed how these parameters change in the setting of acute blood loss have long assumed that hypotension and other signs and symptoms of hemorrhagic shock mark the beginning of circulator...

Страница 22: ...Figure 14 o Before intravascular volume loss 95 of CRI values were above 0 75 and 98 of CRI values were above 0 7 o Before intravascular volume loss the mean CRI value was 0 9 95 confidence 0 89 to 0...

Страница 23: ...on CRI estimates 95 confidence Linear mixed model analysis with CRI as the outcome and intravascular volume changes as the predictor support the conclusion that CRI trends with changes in intravascul...

Страница 24: ...ms CRI estimates are based on reference data collected in ideal conditions Progressive declines to low CRI values indicate measurements based on subjects in clinical studies exhibiting symptoms of inc...

Отзывы: