Flann Microwave SD5902 Скачать руководство пользователя страница 7

Flann Microwave 

 

 SD5902 Waveguide Switch Driver 

SD5902 

Doc No ER15  Version 1.3  

 June 2020

 

 

INSTALLATION 

Interconnection of System Components 

 
The SD5902 Waveguide Switch Driver may be connected to other GPIB instruments using standard 
IEEE488 (GPIB) cables.  Users should pay particular attention to the cable length limitations: total length 
less than 20 m (60 ft), maximum 2 m between GPIB devices. 
 
Waveguide Switches may be connected to the Waveguide Switch Driver via the 2 m (6ft) cables supplied. 
Alternatively, any length of cable up to 5 m (16 ft) may be supplied on request or made up by the user. 
Table 1 shows the wiring details for the 15-way D-type connector 
 

Pin No. 

Function 

1 Large 

Motor 

2 Motor 

Phase 

3 Motor 

Common 

4 Motor 

Phase 

5 Motor 

Phase 

6 Motor 

Phase 

7 Motor 

Phase 

8 Logic 

0v 

9 LNT 

(Precision 

Mode) 

10 Not 

Used 

11 Logic 

+5v 

12 

Output 1 (P1-P2, P3-P4) 

13 

Output 2 (P1-P3) 

14 

Output 3 (P2-P3, P1-P4) 

15 

Output 4 (P2-P4) 

 

Table 1: 15-way D-Type Connections 

 
Use connectors with a casing that incorporates earth terminals and connect the cable screen at both ends to 
ensure earthing integrity.   

Failure to do so may allow EMI radiation to exceed acceptable levels when 

the motor is running. 

  
NOTE: All pins, other than Pin 10, must be connected regardless of the type of switch in use. 
 

Care of Microwave Flanges 

 
The microwave flange surfaces require careful handling to ensure consistent results and long life.  In 
particular, avoid scratching, over-tightening the screws, or leaving the flange surfaces unprotected. Always 
cover the flanges with the protective covers supplied with the switch to prevent ingress of dust, etc., when 
not in use. 
 

WARNING:

 Any grit or protrusions on flange surfaces will cause damage to the mating flange. Before 

mating flanges, each surface should be cleaned with alcohol to remove any grease or particles which may 
have accumulated, and inspected to ensure that the flange is undamaged.  Ensure that debris from cleaning 
materials is not deposited into the waveguide apertures during the cleaning process. 
 

Do not, at any time, insert objects into the waveguide ports as the slightest damage to the inner 
surfaces will impair the microwave performance of the switch. 

 

Содержание SD5902

Страница 1: ...SD5902 Waveguide Switch Driver OPERATION AND MAINTENANCE MANUAL Version 1 3 June 2020...

Страница 2: ...Flann Microwave SD5902 Waveguide Switch Driver SD5902 Doc No ER15 Version 1 3 June 2020 2...

Страница 3: ...COMPONENTS 7 CARE OF MICROWAVE FLANGES 7 USB DRIVER INSTALLATION 8 MANUAL OPERATION 10 FRONT PANEL DESCRIPTION 10 MANUAL CONTROL 10 ERROR INDICATORS 10 TEMPERATURE WARNING 11 REMOTE OPERATION 12 COMP...

Страница 4: ...te the warranty and may result in damage to the product DESIGN CHANGES Flann reserves the right to make changes in the design of its products without reference and without incurring any obligation to...

Страница 5: ...P2 P4 Figure 1 Precision Waveguide Switch Rotor Positions Series 333 switches in conjunction with the SD5902 allow the user to choose between Precision Mode and Speed Mode There are two alternative r...

Страница 6: ...ent must be made inoperative and secured against any unintended operation If this instrument is to be energised via an autotransformer NOT RECOMMENDED for voltage reduction make sure that the common t...

Страница 7: ...that incorporates earth terminals and connect the cable screen at both ends to ensure earthing integrity Failure to do so may allow EMI radiation to exceed acceptable levels when the motor is running...

Страница 8: ...ter and connect a USB cable between the switch driver and computer Power up the switch driver The computer should identify the switch driver and open the Found New Hardware Wizard 2 Select No not this...

Страница 9: ...most systems it will be allocated COM7 or higher It is possible to verify which port has been allocated to the switch driver by checking the Windows Device Manager Firmware Upgrades Users will be able...

Страница 10: ...tween detent positions and that consequently none of the LED position indicators will light If this is the case press any one button and the switch will drive to the selected position There are two mo...

Страница 11: ...rnal temperature of the Waveguide Switch Driver exceeds 55 C the instrument will stop functioning In this state the green and red front panel LEDs will flash alternately and the Temperature Error bit...

Страница 12: ...H1 Acceptor handshake Complete capability T2 Talker Basic talker serial poll no talk only mode no unaddressed if MLA L2 Listener Basic listener no listen only mode no unaddressed if MTA LE0 Extended l...

Страница 13: ...hen a Service Request will be generated upon detection of an error SRQ ON READY If the SR LED illuminates on power on then a Service Request will be generated when the status changes from BUSY to READ...

Страница 14: ...mat a b as shown below if no other query command is present in the command string USB None Front Panel Button 3 in the Instrument A group Command A4 Function Move Switch A to Position 4 Response GPIB...

Страница 15: ...d is present in the command string USB None Front Panel Holding down button 2 in the Instrument B group at power up to cause the light above to extinguish See GPIB Address and Instrument Settings on p...

Страница 16: ...o the current setting of switch A and b is the numerical value assigned to the current setting of switch B See Figure 1 on page 5 This message is retained for compatibility with earlier switch driver...

Страница 17: ...Move Waveguide Switch B to Position 1 B2 Move Waveguide Switch B to Position 2 B3 Move Waveguide Switch B to Position 3 B4 Move Waveguide Switch B to Position 4 B Respond with Waveguide Switch B Posi...

Страница 18: ...cts should be used This unit contains no user serviceable parts In the event of a system malfunction the following steps are suggested Check Line power is available and the power switch is illuminated...

Страница 19: ...nic Equipment WEEE Regulations Flann Microwave is registered with the United Kingdom Environment Agency as a supplier of electrical and electronic equipment and makes the required declarations in acco...

Отзывы: