ENGLISH
- 17
ACCESSORIES
Open the nebuliser by turning the upper part (G1) anticlockwise, re-
move the nozzle (G2) from the upper part (G1) by pressing as shown
by the 2 arrows in the “Connection diagram” in point G.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the acces-
sories, choosing one of the methods described below.
method A:
Sanitise accessories
F-G.1-G.2-G.3-G.4-H-I-L-M
under potable hot water (approximately 40°C) with a gentle, non
abrasive dish detergent.
method B:
Sanitise accessories
F-G.1-G.2-G.3-G.4-H-I-L-M
in the dishwasher with a hot cycle.
method C:
Sanitise accessories
F-G.1-G.2-G.3-G.4-H-I-L-M
by immersing them in a solution of 50% water and 50% white
vinegar, then rinse thoroughly under potable hot water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION paragraph.
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the methods described below.
method A:
Accessories
G.1-G.2-G.3-G.4-H-I-L-M
can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlorite) specific for disinfecting, which is
available in any pharmacy.
Implementation
:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable water and disinfectant, according to the
proportions indicated on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air bubbles on the parts. Leave the parts
immersed for the amount of time indicated on the packaging of the disinfectant associated with the concentration chosen for
the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
method B:
Sanitise the accessories
G.1-G.2-G.3-G.4-H-I-L-M
by boiling them in water for 10 minutes; use demineralised
or distilled water to prevent calcium deposits.
method C:
Sanitise the accessories
G.1-G.2-G.3-G.4-H-I-L-M
with a hot steam steriliser for baby-bottle (not the microwave
type). Perform the process faithfully following the instructions of the steriliser. To ensure that the disinfection is effective, choose
a steriliser with an operating cycle of at least 6 minutes.
At the end of each use store the device complete with accessories in a dry place away from dust.
Switch off the device before any cleaning procedure and unplug the power cable from the socket.
DEVICE AND TUBING EXTERIOR
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any sort).
CLEANING SANITISATION DISINFECTION
Содержание Wi.Neb P0209EM
Страница 23: ...ITALIANO 23...