background image

13

1.  Unless you have used the device before and have applied personal settings, we recommend using the frequency 

adjustment knob (A2) setting it on “stimulation” mode. We also rec-
ommend using the intensity adjustment knob (A4) and turning it 
clockwise (minimum intensity).

2.  Switch on the breast pump by using the switch (A1) located on the 

left-hand side of the control panel of the main body, setting it to “I”: 
the Latty breast pump will start.

3.  Gently turn the intensity adjustment knob (A4) anti-clockwise until 

the effects of suction on the breast and nipple are noticeable but 
not painful.
Normally, we recommend the stimulation stage to last at least 2 
minutes: during this time, you can use the frequency adjustment 
knob (A2) to change the stimulation rhythm from 120 to 80 suction 
cycles per minute.
The setting of the stimulation frequency may vary from woman to 
woman, therefore, the user is advised to try various frequency in-
tervals to find the one that is best for their breasts.

4.  After the stimulation stage, or when the milk starts being expressed, further turn the frequency adjustment knob (A2) 

clockwise to switch to suction mode: the frequency of suction cycles decreases further and the kit’s suction power 
increases. At this point, after setting the most comfort-
able frequency or the one that results in the best expres-
sion (visible through the shield that has been applied to 
the breast), you can further use the intensity adjustment 
knob (A4) so that the applied suction results in proper 
expression of milk.

Warning: In the event the intensity level set should cause 
discomfort or pain, turn the intensity adjustment knob (A4) 
clockwise, thus reducing the force applied to the nipple/
breast.

It is possible to alternate milk expression between the 
breasts or, alternatively, milk can be expressed from both 
breasts by using the double expression kit, which may be 
purchased separately. (Full Twins kit code ACO0021).

5.  At the end of expression, switch the breast pump off. Set the switch (A1) to “0”.

6.  Immediately after each use, thoroughly clean all the kit’s components, as described in the paragraph “Cleaning and 

maintenance”.

To know more

•  Pushing the nipple shield too hard might affect the flow of milk.
•  Do not express milk with too high and uncomfortable vacuum (painful). The pain, which may be a symptom of 

breast and nipple trauma, might decrease the milk flow.

•  Do not try to remove the breast shield during expression. Switch the breast pump off and use a finger to detach the 

shield from the breast; then remove the breast shield.

•  If expression is not comfortable or painful, switch the breast pump off, use a finger to detach the shield from the 

breast and, lastly, remove the breast shield.

•  Do not shorten the tube between the breast shield and the breast pump.
•  Do not use the Latty breast pump sets with an incompatible breast pump.
•  Do not use the Latty breast pump with incompatible breast pump sets.

Wash hands thoroughly with soap and water before touching the breast pump kit(s) 
and the breast. Avoid touching the inside of the containers or caps. Wash the breast with 
water or with suitable wipes.

After assembling the breast pump as indicated in paragraph “Preparing the breast 
pump”, sit in a comfortable and relaxed position. Apply the shield tightly on the breast, 
ensuring the nipple is centred in the shield’s inner cavity.

USE OF THE BREAST PUMP

(A1)

(A2)

(A4)

ST

M

IL

Содержание Latty

Страница 1: ...IJZING PL INSTRUKCJA OBS UGI AR Latty Schema di collegamento Assembly diagram Sch ma de montage Croquis de conexiones Anschlussschema Verbindingsschema Schemacie po cze A B C C8 C8 1 C8 2 C8 3 C8 4 C8...

Страница 2: ......

Страница 3: ...ome ad esempio il ritorno all attivit lavorativa Controindicazioni Non esistono controindicazioni note all utilizzo del tiralatte Latty Descrizione del prodotto Il tiralatte Latty dotato di due regola...

Страница 4: ...e Prima dell uso rimuovere ed eliminare eventuali sacchetti in plastica e tutti gli elementi facenti parte della confezione del prodotto e tenerli fuori dalla portata dei bambini questi potrebbero ess...

Страница 5: ...orrente Non ostruite o introducete oggetti nel filtro e nella sua relativa sede nell apparecchio Non ostruite mai le fessure di aerazione posizionate su entrambe i lati dell apparecchio Controllare ch...

Страница 6: ...orpo principale A ed all attacco tubo sul polmone C8 5 7 Solo per taglia M disponibile un adattatore morbido in silicone per offrire maggior comfort solo per la taglia media 27mm In tale eventualit ut...

Страница 7: ...stimolazione pu variare da don na a donna e quindi si invita l utente a provare diversi intervalli di frequenza per trovare quello che meglio si adatta al proprio seno 4 Dopo la fase di stimolazione o...

Страница 8: ...lembi di chiusura centrali Nota per rimuovere il tubo in silicone B staccare prima il sistema polmone C8 dal tappo porta coppette C5 in modo da evitare rovesciamenti Afferratelo quindi all estremit a...

Страница 9: ...completamente asciutti assemblare nuovamente il kit ti ralatte Se dopo aver verificato le condizioni sopradescritte l apparecchio non dovesse ancora funzionare correttamente consigliamo di rivolgervi...

Страница 10: ...reless potrebbero interferire con il funzionamento dei dispositivi elettromedicali Per ulteriori informazioni visitate il sito internet www flaemnuova it Il Dispositivo Medico potrebbe essere suscetti...

Страница 11: ...rn for any reason for example if she is going back to work Contraindications There are no known contraindications to the use of the Latty breast pump Product description The Latty breast pump is fitte...

Страница 12: ...operty Carefully follow the instructions in this manual to assure proper operation and to extend the service life of the device Do not expose the device and its components to direct sunlight rain and...

Страница 13: ...e Make sure there is no material obstructing the air vents before each use Do not put any objects in the air vents The device is fitted with an easy to inspect safety fuse in case of fault Before doin...

Страница 14: ...ing housing A5 on the main body A and to the tube connection on the diaphragm C8 5 7 Only for size M a soft silicone adapter is available that offers more comfort only for size M 27 mm In that case us...

Страница 15: ...t or pain turn the intensity adjustment knob A4 clockwise thus reducing the force applied to the nipple breast It is possible to alternate milk expression between the breasts or alternatively milk can...

Страница 16: ...entral closing flaps Note to remove the silicone tube B first disconnect the diaphragm system C8 from the shield cap C5 to avoid overturning Then grip it from the end that is attached to the upper dia...

Страница 17: ...e When the components are completely dry assemble the breast pump kit once more If after checking the above described conditions the device is still not working properly we recommend taking it to your...

Страница 18: ...communications devices mobile telephones or wireless connections might interfere with the operation of electro medical devices For additional information please go to www flaem it The Medical Device m...

Страница 19: ...ue ce soit telle que le retour au travail Contre indications Il n existe aucune contre indication connue l utilisation du tire lait Latty Description du produit Le tire lait Latty est quip de deux r g...

Страница 20: ...ssance de ce manuel Avant l utilisation retirer et liminer tous les sacs en plastique et tous les l ments faisant partie de l emballage du produit et les conserver hors de la port e des enfants ceux c...

Страница 21: ...t branch une prise de courant Ne pas obstruer ni introduire d objets dans le filtre ni dans son logement pr vu dans l appareil Ne jamais obstruer les trous d a ration positionn s des deux c t s de l a...

Страница 22: ...du tuyau A5 du corps principal A et au raccord du tuyau sur le poumon C8 5 7 Uniquement pour la taille M un adaptateur en silicone souple est disponible pour un plus grand confort uniquement pour la t...

Страница 23: ...n d sagr able ou douloureuse utiliser le bouton de r glage de l intensit A4 en le tournant dans le sens horaire r duisant ainsi la force appliqu e sur le mamelon sein Il est possible d alterner le tir...

Страница 24: ...s de fermeture centraux Remarque pour enlever le tube en silicone B d tacher d abord le syst me poumon C8 du bouchon porte coquille C5 afin d viter tout renversement Puis le saisir l extr mit accroch...

Страница 25: ...embler le kit du tire lait Si apr s les essais d crits ci dessus l appareil ne fonctionne toujours pas correctement il faut contacter votre revendeur de confiance ou le centre d assistance agr FLAEM l...

Страница 26: ...communication mobiles ou portables RF t l phones portables ou connexions sans fil peuvent interf rer avec le fonctionnement des dispositifs lectro m dicaux Pour plus d informations visitez le site We...

Страница 27: ...icaciones No existen contraindicaciones conocidas relacionadas con el uso del extractor de leche Latty Descripci n del producto El extractor de leche Latty dispone de dos reguladores que permiten vari...

Страница 28: ...e do este manual Antes del uso quite y elimine las bolsas de pl stico y todos los elementos que formen parte del embalaje del producto y mant ngalos fuera del alcance de los ni os estos podr an ser fu...

Страница 29: ...ca objetos en el filtro ni en su alojamiento No obstruya nunca las ranuras de ventilaci n situados en los dos lados del aparato Controle que no haya material que obstruya las ranuras de ventilaci n an...

Страница 30: ...el cuerpo principal A y el enganche del tubo en el diafragma C8 5 7 Solo para el tama o M est disponible un adaptador suave de silicona para ofrecer un mayor confort solo para el tama o mediano 27mm E...

Страница 31: ...egulador de intensidad A4 gir ndolo a la derecha reduciendo as la fuerza aplica da en el pez n seno Se puede alternar la extracci n de la leche de los dos se nos o como alternativa se puede extraer la...

Страница 32: ...para quitar el tubo de silicona B desmonte primero el sistema de diafragma C8 del tap n de soporte de las copas C5 para evitar vuelcos Despu s af rrelo por el extremo enganchado en el diafragma superi...

Страница 33: ...lo Limpieza y mantenimiento Cuando los componentes est n completamente secos monte de nuevo el kit del ex tractor de leche Si despu s de haber constatado las condiciones arriba expuestas el equipo sig...

Страница 34: ...fonos m viles o conexiones inal mbricas podr an interferir con el funcionamiento de los equipos electrom dicos Para m s informaci n visite el sito internet www flaemnuova it El Aparato M dico puede s...

Страница 35: ...geborenen aus einem beliebigen Grund zu trennen wie zum Beispiel die R ckkehr zur Arbeitst tigkeit Gegenanzeigen Es sind keine bekannten Gegenanzeigen f r die Verwendung der Milchpumpe Latty vorhanden...

Страница 36: ...iese f r Kinder unzug nglich aufbewahren diese k nnen eine Gefahrenquelle sein Einige Ger teteile sind sehr klein und k nnten von Kindern verschluckt werden Bewahren Sie das Ger t daher au er halb der...

Страница 37: ...hlitze auf beiden Seiten des Ger tes Vor jeder Benutzung sicherstellen dass die Luftschlitze nicht verstopft sind Keine Fremdk rper in die Luftschlitze einf gen Das Ger t ist mit einer Sicherung verse...

Страница 38: ...erbindungsschlauches A5 mit dem Hauptger t A und dem Schlauchanschluss am Speicher C8 5 verbinden 7 Nur f r Gr e M F r einen besseren Komfort ist nur f r die Gr e M 27 mm ein weicher Silikonadapter er...

Страница 39: ...e Brust warze Brust angewendete Kraft reduzieren Es ist m glich die Milch abwechselnd aus beiden Br sten abzupumpen oder alternativ dazu die Milch aus beiden Br sten durch Verwendung des Doppelkits ab...

Страница 40: ...mittleren Verschlussr nder einwirken Hinweis Zum Entfernen des Silikonschlauchs B zuerst das Speichersystem C8 vom Halterungsverschluss der Brust h tchen C5 entfernen um ein Umkippen zu vermeiden Dan...

Страница 41: ...montieren Wenn nach der Pr fung all der oben beschriebenen Bedingungen das Ger t dennoch nicht korrekt funktionieren sollte raten wir Ihnen sich an den Wiederverk ufer oder an das am N chsten liegend...

Страница 42: ...e oder Wireless Anschl sse k nnten den Betrieb der elektromedizinischen Ger te st ren F r weitere Informationen besuchen Sie die Homepage www flaemnuova it Das Medizinger t k nnte wenn andere Vorricht...

Страница 43: ...baby is gescheiden bijvoorbeeld om weer te gaan werken Contra indicaties Er zijn geen contra indicaties bekend bij het gebruik van het kolfapparaat Latty Beschrijving van het product Het kolfapparaat...

Страница 44: ...cht van een volwas sene worden gebruikt die deze handleiding aandachtig heeft gelezen V r het gebruik moet u eventuele plastic zakjes en alle verpakkingselementen van het product verwijderen daar ze e...

Страница 45: ...et met een stopcontact is aangesloten De filter en de filterzitting van het apparaat niet afdekken of er voorwerpen insteken Belemmer de ventilatieopeningen aan de zijkanten van het apparaat niet Cont...

Страница 46: ...de balg C8 5 7 Alleen voor maat M er is een zachte adapter in silicone beschikbaar om meer comfort te bieden alleen voor de maat medium 27 mm Als u de borstschelp maat 30 mm C4 gebruikt de adapter in...

Страница 47: ...fwel kunt u de melk van beide borsten tegelijk afkolven met behulp van de kit voor dubbele extractie die u apart moet kopen Com plete kit Twins code ACO0021 5 Na het afkolven schakelt u het kolfappara...

Страница 48: ...pmerking om de slang in silicone B weg te nemen maakt u eerst het balgsysteem C8 van het borstschelpdeksel los C5 zodat het niet omvalt Neem daarna het uiteinde vast dat aan de bovenste balg C8 3 is v...

Страница 49: ...Wanneer de componenten volledig droog zijn assembleert u de borstkolfset opnieuw Als het toestel nog steeds niet correct werkt nadat het bovenstaande werd gecontroleerd raden wij u aan om u te wenden...

Страница 50: ...erking van elektromedische apparaten Surf naar de website www flaemnuova it voor meer informatie Het Medisch Hulpmiddel kan gevoelig zijn voor elektromagnetische interferentie als andere apparaten aan...

Страница 51: ...dku nadmiaru pokarmu lub konieczno ci pozostawienia noworodka z jakiegokolwiek powodu takiego jak na przyk ad powr t do pracy Przeciwwskazania Nie s znane przeciwwskazania do stosowania laktatora Latt...

Страница 52: ...dzorem osoby doros ej kt ra zapozna a si z niniejsz instrukcj Przed u yciem urz dzenia usun wszelkie woreczki plastikowe i wszystkie elementy kt re s cz ci opakowania produktu i przechowywa je poza za...

Страница 53: ...cych si po obu stronach urz dzenia Przed ka dym u yciem nale y sprawdzi czy otwory wentylacyjne nie s zatkane Nie wk ada adnych przedmiot w do otwor w wentylacyjnych Urz dzenie jest wyposa one w bezp...

Страница 54: ...a rozmiaru M dost pny jest mi kki silikonowy reduktor zapewniaj cy wi kszy komfort tylko dla rednich rozmiar w 27 mm W takim przypadku u ywaj c lejka o rednicy 30 mm C4 na o y na niego silikonowy redu...

Страница 55: ...y powoduje uczucie dys komfortu lub b l obr ci pokr t o do regulacji intensywno ci A4 w prawo zmniejszaj c w ten spos b si oddzia uj c na sutek pier Mo liwe jest naprzemienne pobieranie mleka z dw ch...

Страница 56: ...jek C5 aby unikn przechylenia Nast pnie przytrzyma j za koniec przy czony do g rnej komory C8 3 i lekko poci gn ODKA ANIE Przed i po ka dym u yciu akcesoria nale y odkazi w jeden z ni ej opisanych spo...

Страница 57: ...wdzeniu wy ej opisanych warunk w urz dzenie nadal nie pracuje prawid owo nale y skontaktowa si z odsprzedawc lub z najbli szym autoryzowanym serwisem FLAEM Metoda B Zdezynfekowa akcesoria C1 C 2 C 3 C...

Страница 58: ...tliwo radiow telefonykom rkowelubpo czeniabezprzewodowe mo ezak ca prac sprz tuelektromedycznego Aby uzyska wi cej informacji zapraszamy na stron internetow www flaemnuova it Wyr b medyczny mo e by po...

Страница 59: ...1 H IG N B H 78 I H 1 J Latty I 2 8 0 2 4 3 Q N 1 1 Latty I 9 9 9 1 7 69 1 M 2 78 7 0 3 2 2 1 6 M3 8 3 3 D 2 9 7 M Q 3 H J 6 0 RH J M Q 3 1 M Q H J M Q MS B 1 M3 2 78 Q 78 7 N A 23 T U A1 3 A U A2 3...

Страница 60: ...78 4 9 1 78 B 4 8 B B 6 D z 7 8 B P z 6 0 6 6 8 K 0 7 1 5 z 8 78 B B 6 z 0 6 Q b 6 B C 78 49 8 1 1 B 9 z 1 0 MFLAEM 1 0 6 3 6 1 Q 6 6 N 78 MT 9 1B H z 3 6 1B 1 1 7Q T 9 n T 9 0 9 Q 700 3n 6 Q 3 z 0 9...

Страница 61: ...0 1 6 7 9 0 2 b B IB 6 7 M 6 z 3 9 1 1B 1 6 3G 6 1 0 N E6 3G 6 1 3 1 Q9 A0 z 6 3G 6 78 78 MK 78 3 3 Q 6 H 1B 1 6 3G 6 1 0 4 0 B z 6 3 H 1B 1 0 M 78 Q9 G 1 0 D3 B 2 z 2 2 1 8 K8 6 b 3 1B 1 5 M 3 5 3 6...

Страница 62: ...J 3 1 A4r B 6 3 H A 7 8 M G B 0 0 78 S H R4 6 7 M H 78 K 3 1 A4r 3 I x J 1 MN 1 B M1 1 H J 6 1 1 1 7 M J 1 1 H ACO0021 6 Twins 3r 69 6 4 A1r T 7 I 7 nB MJ 78 5 0 2 6 0 8 2 6 C 7 6 9 C 8 78 1 H K8 Q 6...

Страница 63: ...1 78 C4 B C3 B C2r H 3 3 Q9 b5 6 H 5 C8 5r 78 3 9 Ar 7 6 A5r 3 0 Br 3 9 3 6 9 6 M C4r 30 3 Q9 b5 M 78 27r 3 b 3 8 B 8 1 6 8 R M b 3 8 7 4 B 1 5 2 Mb5 8 8 M7 9 7 3 60 M4 C6r 1 6 4 5 3 6 3 K9 0 B A6r 7...

Страница 64: ...N 1 6 78 70 b 1 3 6 7 98 M 6 Q 7 B Q 3 I C nr 3 03 7 2 C 7 R 3 78 H B C B N 1 9 C 698 78 D I I 2 I B M 8 G 1 M4 8 9 C 1 5 0 MD 9 6 0 6 0 N H G Flaem 1 0 7 B 6 2 4 9 7 EK 10 78 n K 3 1 D F G H J J J J...

Страница 65: ...c S 6 0 6 1 C5r b5 6 3n 6 3 1 C9r J 9 N 7 I 2 7 98 n C b 1 0 78 4 0 4 H C5r b5 6 3n 1 C8r 69 B 7 M Br 1 3 R C3 4 K 3 1 C8 3r 0 7 9 C 3 S Q I 3 M 6 0 6 40 r H D 9 1 D F G H J J J J K I 3 A 3 nr 3G C C...

Страница 66: ...64...

Страница 67: ...12 A r 26 R1 2 100 3 B R7 BF 3 G M M r S 2 1UEN 60 60601r 7 3 Q K8 B K 6 B 1 3 3 4 9 1B H 3 8 7 M 3 Q 3 3 K 0 8 M H 78 M 9 H 3 3 G J 6 B 2 G MS B B 2 3 Q 3 0 9 4 8 Bx 2 6 9 B C r 0 G B 9 B S B 7 3 1...

Страница 68: ...sistenzaTecnica Autorizzato nel caso di operativit della garanzia Nel caso in cui il Centro di Assistenza Tecnica Autorizzato dovesse riscontrare che l apparecchio esente da difetti e o risulta funzio...

Отзывы: