CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket.
DEVICE (A) AND TUBING EXTERIOR (B)
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any
sort).
ACCESSORIES
Open the nebuliser by turning the upper part (C1.3) anticlockwise, remove the nozzle
(C1.2) from the upper part (C1.3) by pressing as shown by the 2 arrows in the "Connection
diagram" in point C1.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of
the methods described below.
method A:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
under potable hot
water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
in the dishwasher
with a hot cycle.
method C:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
by immersing
them in a solution of 50% water and 50% white vinegar, then rinse thoroughly under
potable hot water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below.
method A:
Accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C5-C6
can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
USE OF NEBULIZATION MANUAL CONTROL
To achieve continuous nebulisation action you should not use the manual nebulisation
control (C6), especially in the case of children or persons with reduced physical, sensory, or
mental capabilities. The manual nebulisation control is useful for limiting dispersion of the
medication in the surrounding environment.
To start nebulizing close
with a finger the hole of the
nebulizer manual control
(C6) and breathe in gently;
we recommend to hold your
breath for a moment so that
the inhaled aerosol droplets
can be deposited,
meanwhile, to disable nebu-
lizing, remove your finger
from the hole of the nebu-
lizer manual control to avoid
the waste of drug, optimiz-
ing its acquisition. Then ex-
hale slowly.
4
5
Содержание JN33P00
Страница 2: ......
Страница 52: ...FLAEM FLAEM F400 400 F700 700 F1000 1000 F2000 2000 a b 93 42 CE 2 50...
Страница 53: ...1 A6 A5 2 C1 2 C1 3 2 C1 C1 4 C1 3 C1 C1 1 C1 1 3 4 5 1 6 B RF7 DUAL SPEED RF7 Dual Speed C1 4 MAX MIN C1 4 51...
Страница 56: ...RF7 Dual Speed ACO461P ACO462P 1 ACO35 2 ACO439 ACO463P FLAEM 54...
Страница 57: ...T V CE 93 42 CEE II BF EN 10993 1 93 42 EEC CE 1907 2006 IP21 IP21 12 EN60601 1 2 www flaemnuova it Flaem 55...
Страница 59: ...6 Neb aid P0603EM F400 www flaem it 7 RF7 C1 C1 1 C1 2 C1 3 C1 4 C2 C3 SoftTouch C4 C5 C6 C 57...
Страница 60: ...SoftTouch SoftTouch C1 4 MAX MIN C1 4 RF7 Dual Speed Dual Speed RF7 4 A1 5 6 B 58...
Страница 61: ...F1000 F700 700 F400 400 F2000 2000 1000 CE 42 93 A5 A6 1 C1 3 C1 2 2 C1 3 C1 4 C1 C1 1 C1 C1 3 3 59...
Страница 62: ...C1 1 C1 2 C1 3 C1 4 C2 C4 C3 C5 C6 B 10 C1 1 C1 2 C1 3 C1 4 C2 C4 C3 C5 C6 C 6 A3 60...
Страница 64: ...T V CE CEE 93 42 BF EN10993 1 EEC 93 42 1907 2006 CE IP21 IP21 12 EN 60 60601 1 2 www flaemnuova it 62...
Страница 65: ...ACO461P RF7 Dual Speed ACO462P ACO35 1 ACO439P ACO463P 63...
Страница 67: ......