10
11
SYMBOLS
TÜV approval
EC Marking medical ref. Dir 93/42 EEC and subse-
quent updates
Device serial number
Class II device
Manufacturer
Attention: check the instructions for use
Type BF applied part
“OFF” for part of Equipment
Alternating current
“ON” for part of equipment
Keep dry
In compliance with: European Standard EN 10993-
1 ‘’Biological Evaluation of medical devices’’ and
European Directive 93/42/EEC “Medical Devices.”
Phthalate-free. In compliance with: Reg. (EC) no.
1907/2006
Risk: electrocution.
Consequence: Death.
Using the device while taking a bath or shower is
prohibited
Minimum and maximum room temperature
Minimum and maximum air moisture
Minimum and maximum atmospheric pressure
TROUBLESHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
CAUSE
SOLUTION
The device does not
work
The power cable has not been
correctly inserted into the socket of
the device or the power socket
Correctly insert the power cable in sockets.
The device does not
nebulize or
nebulizes
insufficiently.
The medication has not been
inserted in the nebulizer
Pour the right amount of medication in the nebulizer
The nebulizer has not been
properly fitted.
Disassemble the nebulizer and reassemble it correctly as
shown in the connection diagram on the cover.
The nebulizer nozzle is clogged
Disassemble the nebulizer, remove the nozzle and
perform cleaning operations.
Failure to remove medication deposits from the nebulizer
affects its efficiency and operation.
Strictly comply with the instructions
contained in the “CLEANING, SANITISATION,
DISINFECTION” chapter
The air pipe is not correctly con-
nected to the device
Ensure the device’s accessories are properly connected to
the air intake (see connection diagram on the cover).
The air pipe is bent, damaged or
twisted
Unwind the pipe and ensure it is not crushed or
punctured. If necessary replace it.
The air filter is dirty
Replace the filter
The apparatus is
noisier than usual
The Filter is not correctly
inserted in its housing
Fully and properly insert the Filter in its housing
If after verifying the above mentioned conditions the device should not operate properly, contact your trusted retailer
or the nearest authorized FLAEM service center.
Содержание CO03P00
Страница 43: ...41 FLAEM T V CE 93 42 CEE II BF EN 10993 1 93 42 EEC CE 1907 2006...
Страница 45: ...43 P0406EM F700 www flaem it 6 0 0...
Страница 46: ...44 1000 F1000 700 F700 400 F400 2000 F2000 CE 42 93...
Страница 49: ...47 T V CE CEE 93 42 BF EN10993 1 EEC 93 42 1907 2006 CE...
Страница 50: ...48 EN6060601 1 2 www flaemnuova it CE 2002 96 CE 2002 96...
Страница 51: ......