NASAL WASH RHINO CLEAR®
Before each use, wash your hands thoroughly and sanitize RHINO CLEAR® as described in section
“CLEANING SANITIZATION DISINFECTION”. The NASAL WASH SYSTEM and accessories are for personal
use only to avoid potential risks of transmitting infection. During the application it is recommended to be
adequately protected against dripping.
1) Insert the atomizer (C5.3) into the nebulization chamber (C5.4.1).
2) Place the solution (max 10 ml) in the nebulizing chamber (C5.4.1).
3) Apply the separator (C5.2) verifying that the arrow on the shower body is aligned
with the separator (Pic. B). Then, insert the correct adapter for your age, or the most
appropriate to fit the size of your nostrils (C5.1A/C5.1B/C5.1C).
4) Connect the RHINO CLEAR® device through the tube (B) and the fitting (C6).
5) Turn on the device using the switch (A1).
6) Place the device close to the nose and gently insert the adapter nosepiece into one
nostril.
7) During treatment it is preferable to hold the other nostril closed, where the shower is not applied, by applying
light finger pressure.
8) Run the shower holding down the manual dispensing button (C5.5) during the inhalation phase and release
it before exhaling; at this stage, to help expel the mucus, blow gently through the nostril treated directly into
the RHINO CLEAR®.
9) Repeat from Step 6 to Step 8 for the other nostril, trying to evenly distribute the solution between the two
nostrils.
10) When used on children or other dependent people, or in any case when there are problems in coordinating
the various phases of paragraph 8), we recommend to run the application alternating the nostrils about 20
seconds each, with pauses required to expel the mucus.
11) The operations described in the preceding paragraphs is useful for nasal cleaning using physiological
solutions. The treatment with drugs is advisable after a first cleaning with physiological solution in order to
achieve a more effective distribution of the drug on the nasal mucosa.
12) Once the application is complete, switch off the device and clean it, following the instructions provided in the
“CLEANING SANITISATION DISINFECTION” paragraph.
CLEANING SANITISATION DISINFECTION
Switch the device off and disconnect it from the power supply before proceeding with any cleaning operation.
CLEANING THE DEVICE AND THE TUBE’S OUTER SURFACE
Only use a cloth with antibacterial detergent (non-abrasive and with no solvents of any kind)
ACCESSORIES
We recommend personal use of accessories to avoid any risk of spreading the infection.
Disassemble the accessories as described below:
C1
- Open the nebulizer by turning the upper part (C1.3) anticlockwise, remove the nozzle from the upper part
(C1.3) and then disassemble it as shown in the “Assembly diagram” in section C1.2.
C5
- Disassemble it as shown in the “Assembly diagram” in section C5.
SANITISATION AND DISINFECTION
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of the methods described
below.
method A:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5.1A-C5.1B-C5.1C-C5.2-C5.3-C5.4-C5.5-C6
under potable hot water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5.1A-C5.1B-C5.1C-C5.2-C5.3-C5.4-C5.5-C6
in the dishwasher with a hot cycle.
method C:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5.1A-C5.1B-C5.1C-C5.2-C5.3-C5.4-C5.5-C6
by immersing them in a solution of 50% water and 50% white vinegar, then rinse thoroughly under potable hot
water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper towel. Alternatively, dry
them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the methods described be-
C5.4.1
6. Turn on the device using the switch (A1) and then breathe in and exhale deeply.
7. When the treatment is finished, shut off the device and unplug the power cord.
WARNING:
If after the therapy session moisture accumulates in the tube (B), remove the tube from the nebulizer
and dry it using the device’s compressor fan; this operation will prevent mold from forming inside the tube.
Содержание CO03P00
Страница 43: ...41 FLAEM T V CE 93 42 CEE II BF EN 10993 1 93 42 EEC CE 1907 2006...
Страница 45: ...43 P0406EM F700 www flaem it 6 0 0...
Страница 46: ...44 1000 F1000 700 F700 400 F400 2000 F2000 CE 42 93...
Страница 49: ...47 T V CE CEE 93 42 BF EN10993 1 EEC 93 42 1907 2006 CE...
Страница 50: ...48 EN6060601 1 2 www flaemnuova it CE 2002 96 CE 2002 96...
Страница 51: ......