CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket.
DEVICE AND TUBING EXTERIOR
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any
sort).
ACCESSORIES
Open the nebuliser by turning the upper part (C1.3) anticlockwise, remove the nozzle
(C1.2) from the upper part (C1.3) by pressing as shown by the 2 arrows in the "Connection
diagram" in point C1.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of
the methods described below.
method A:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5
under potable hot
water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5
in the dishwasher with
a hot cycle.
method C:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5
by immersing them in
a solution of 50% water and 50% white vinegar, then rinse thoroughly under potable hot
water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below.
method A:
Accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5
can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
SoftTouch
masks have an outer edge made of
soft
biocompatible material
that ensures excellent ad-
herence to the face, and is also equipped with an
innovative Dispersion Limiting Device
. These dis-
tinctive elements that distinguish it allow greater
sedimentation of medication in the patient and also
limit dispersion
.
During the inspiratory phase, the
tab
, acting as a
Dispersion Lim-
iting Device
, bends inwards to-
wards the mask.
During the expiratory phase, the
tab
, acting as a
Dispersion Limit-
ing Device
, bends outwards from
the mask.
SOFTTOUCH FACE MASKS
Soft
biocompatible
material
Dispersion
Limiting Vent
12
Содержание AlphaNeb
Страница 2: ......
Страница 52: ...FLAEM FLAEM F400 400 F700 700 F1000 1000 F2000 2000 a b 2 50...
Страница 53: ...1 A5 2 C1 2 C1 3 2 C1 C1 4 C1 3 C1 C1 1 C1 3 3 4 5 1 6 SOFTTOUCH SoftTouch 51...
Страница 56: ...E FLAEM 54...
Страница 57: ...EN 60601 1 2 EMC www flaemnuova it Flaem T V 93 42 II BF EN 10993 1 93 42 CE 1907 2006 IP21 IP21 12 mm 55...
Страница 59: ...Alpha AlphaNeb Neb 3 0 ODHP ZZZ DHP LW 5 XDO 6SHHG 6RIW 7RXFK 6RIW 7RXFK 57...
Страница 60: ...0 0 K K K K D E 58...
Страница 61: ...5 8 63 5 XDO 6SHHG 0 0 1 59...
Страница 62: ...r r 62 7728 6RIW7RXFK 60...
Страница 63: ...ODHP 5 XDO 6SHHG 2 3 2 3 2 2 3 2 3 61...
Страница 64: ...0 62...
Страница 65: ...7 9 1 Q 3 3 PP 1 0 5 ZZZ DHPQXRYD LW ODHP 63...
Страница 75: ......