Operation
Preparing the device
The device must not be released for operation before all major start-up activities have been
completed (see chapter
Device Check
Prior to starting operation, the following device components must be checked for their correct
function:
• The door seal in the front frame must not be damaged.
• The perforated plates have to have been correctly inserted.
Starting Operation
1. Switch the device on with button
.
2. The display shows the current useable space temperature.
3. Set the target values for the temperature using buttons
and
on the control
panel.
4. Wait until the device has reached the set temperature. The temperature control regulates
the temperature at the set target temperature value.
5. Load the useable space.
Proper loading
To ensure sufficient air circulation and uniform heating of the sam-
ples, do not use more than 70% of the maximum surface area of the
useable space. Bulky objects in the useable space that dissipate
heat may impair heat distribution.