background image

Type 4000

9

and connected to the actuator power module through
a feedback linkage, thus providing feedback from the
actuator/valve. When these two forces are equal, the
balance arm and the spool in the pilot valve is in a
neutral position—the complete unit is in a balanced
position. The pilot valve controls the air, supplied
through port S, to the actuator.

Assume an equilibrium position. An increasing control
pressure deflects the diaphragm 1 to the right, com-
pressing the feedback spring 3. The balance arm 2
moves the spool 7 in the pilot valve 8 furnishing supply
air to the actuator.

With increasing supply air, the linear motion of the ac-
tuator stem rotates the positioner spindle 6. The
spindle and cam 5 rotation forces the lower arm 4 to
the left compressing the feedback spring 3. This mo-
tion continues until the two forces are equal and the
unit is in an equilibrium position.

Maintenance

WARNING

Avoid personal injury from sudden re-
lease of process pressure. Before per-
forming any maintenance operations:

Disconnect any operating lines pro-

viding air pressure or a control signal to
the actuator. Be sure the actuator can-
not suddenly open or close the valve.

Use bypass valves or completely

shut off the process to isolate the valve
from process pressure. Relieve process
pressure on both sides of the valve.
Drain the process media from both
sides of the valve.

Vent actuator loading pressure and

relieve any actuator spring precompres-
sion.

Use lock-out procedures to be sure

that the above measures stay in effect
while you work on the equipment.

Other than the pilot valve, repair of the positioner is by
replacement. To replace the positioner, see the Posi-
tioner Removal and Positioner Replacement sections.
To replace the pilot valve, see Replacing the Pilot
Valve. On units with gauges, to replace the gauges,
see Replacing Gauges.

Troubleshooting

Signal change has no effect on the actuator posi-
tion.

—Check indicator and screw.

—Check air supply to positioner.

—Check input signal to positioner.

—Check diaphragm for damage or leakage.

—Check pilot valve function.

—Check cam for correct setting.

Signal change results in actuator running to end
positions.

—Check coupling between positioner and actuator.

—Check cam position and locking screw.

—Check input signal.

Inaccurate positioning.

—Dirty or worn pilot valve.

—Defective or leaking diaphragm.

—Input signal fluctuates.

—Incorrect sizing of actuator.

—Valve/actuator “stiction”.

—Loose cam.

Replacing the Pilot Valve

Refer to figure 12 to remove the pilot valve for clean-
ing or inspection.

1.  Shut off supply pressure to the positioner.

2.  Remove the positioner front cover.

3.  Remove the screw and carefully lift out the com-
plete pilot valve assembly.

4.  Gently remove the spool from the block and clean
the parts, using methylate cleaner or similar. Blow the
parts dry with compressed air. If the parts show signs
of wear, replace the assembly.

Note

Mixing spool valves and valve bodies
may result in very high bleed rates and
poor performance.

Содержание 4000

Страница 1: ... Replacing the Positioner 10 Parts Ordering 11 Parts List 11 Introduction Scope of Manual This instruction manual provides installation operat ing calibration maintenance and parts ordering in formation for the Type 4000 pneumatic valve Figure 1 Type 4000 Positioner Installed on System 9000 Actuator W7371 IL positioner Refer to separate instruction manuals for information on the actuator control v...

Страница 2: ...tic valve positioner is designed for use with System 9000 actuators The positioner mounts on the actuator and provides the desired valve plug position for a specific input signal The Type 4000 accepts a pneumatic input signal The positioner comes standard with gauge ports and an O ring sealed housing The housing utilizes a unique O ring seal that can be adjusted to a sealed or drained position Eas...

Страница 3: ...ey 1B from the positioner spindle by removing the flathead screw that holds the lever to the spindle Figure 3 Assembling the Type 4000 Positioner to the Adapter Housing ADAPTER HOUSING KEY 1C POSITIONER KEY 1A MOUNTING SCREWS W7369 IL Note Mounting the positioner key 1A may be easier with the adapter housing key 1C removed from the actuator 2 To remove the adapter housing key 1C from the actuator ...

Страница 4: ... USE THESE HOLES FOR SIZE 20 ACTUATOR USE THESE HOLES FOR SIZE 50 AND 80 ACTUATORS W7048 IL 9 Refer to figure 6 Install the feedback bracket key 108 to the top of the power module with two cap screws key 105 and washers key 122 10 If necessary make minor adjustments of the feed back lever arm assembly pin to ensure that the feed back bracket and feedback lever arm assembly do not pinch bind or ben...

Страница 5: ...ticles 40 microns in diameter should suffice for most applications Use of suitable filtration equipment and the es tablishment of a maintenance cycle to monitor its operation is recommended The supply medium must be clean dry air or noncor rosive gas that meets the requirements of ISA Stan dard S7 3 1975 R1981 The supply pressure should not exceed the following limits For the actuator and position...

Страница 6: ...on and necessary safety measures should be used There is no pipe fitting for remote venting The posi tioner exhaust is under the actuator cover Operating Information Front Cover and Indicator Cover The positioner front cover is shown in figure 8 It is secured to the unit with four captive screws and sealed with an O ring The front cover O ring can be looped over notches in the front cover to allow...

Страница 7: ...ure the cam make sure the screw is backed out from the locking nut Tighten the locking nut finger tight then tighten the screw Install and adjust the indi cator and reinstall the cover Indicator Adjustment To adjust the indicator remove the front cover and pull the indicator upwards until it comes off the Allen screw Before installing the indicator make sure that the Allen screw is tight Press the...

Страница 8: ...re 11 Principle of Operation 1 2 3 4 5 6 7 8 A7199 IL If the actuator travel recorded in step 5 is short of the desired travel at 100 input signal rotate the span adjustment down to increase span Tighten the locking screw If the travel recorded in step 5 reaches the de sired travel before the input signal reaches the 100 value rotate the span adjustment up to decrease span Tighten the locking scre...

Страница 9: ...loading pressure and relieve any actuator spring precompres sion Use lock out procedures to be sure that the above measures stay in effect while you work on the equipment Other than the pilot valve repair of the positioner is by replacement To replace the positioner see the Posi tioner Removal and Positioner Replacement sections To replace the pilot valve see Replacing the Pilot Valve On units wit...

Страница 10: ...1 Shut off supply pressure and process lines to the positioner 2 Remove the defective gauge from the adapter housing 3 Coat the threads of the replacement gauge with a sealant Zink plate No 770 or equivalent Figure 13 Diaphragm DIAPHRAGM COVER SCREWS SCREW WASHERS DIAPHRAGM A7201 IL 4 Screw the replacement gauge into the adapter housing gauge connection 5 Apply supply pressure to the positioner Us...

Страница 11: ...ation Cover For 3 4 19 1 mm travel or less P5FV30R 18B2836X012 For greater than 3 4 19 1 mm travel P5FV60R 18B2835X012 3 Instrument Gauge Triple Scale 0 30 psi 2 bar 0 2 MPa 11B4036X012 4 Supply Gauge Triple Scale 0 160 psi 11 bar 1 1 MPa 11B4036X032 Key Description Part Number 5 Output Gauge Triple Scale 0 160 psi 11 bar 1 1 MPa 11B4036X032 Test Connections use when gauges or pipe plugs are not s...

Страница 12: ... 05424 Brazil Singapore 128461 Printed in U S A Fisher Controls International Inc 1998 1999 All Rights Reserved Fisher Fisher Rosemount and Managing The Process Better are marks owned by Fisher Controls International Inc or Fisher Rosemount Systems Inc All other marks are the property of their respective owners ...

Отзывы: