background image

Packing List

The following items are packed in the envelope located

inside the freezer chamber. If any of the following items

are not present, report the missing item to your local

Fisher representative.

1. Instruction Manual

3. Door Lock Key

4. Power Switch Key

5. Chart Recorder Instructions

(Catalog No.ʼs with “R”)

Selecting a Location

Choose a location for the freezer that will provide

at least

three inches of clearance

between the cabinet and any

adjacent vertical surface at the sides and rear.

Appropriate electrical power must be available. Locate

the freezer within 6 feet of the power outlet so that no

extension cord is required.

An installation instructions form is attached to the

front of the unit. After installation, confirm that the

unit was installed per these instructions, that the

required information was completed and that the form

was faxed to the specified number.

Installing and Leveling Casters

Freezers are shipped with four unattached threaded cast-

ers. Casters screw directly into the weld nut in each cor-

ner of the base of the unit. Use the wrench provided to

thread the casters completely into the base of the unit.

Back the casters in or out until the unit is level and resting

on all four casters. See figure 1.

U

NPACKING AND

I

NSTALLATION

Freezer Base 

Thread casters directly

into base at four corners.

Tighten using
wrench provided.

Note

The freezer must be level in order to

provide adequate condensation

drainage as well as proper door

alignment and operation. The freez-

er should be in its final operating

location and set so that it is firmly

positioned on the floor.

5

Figure 1

Содержание Isotemp Plus 13-986-142

Страница 1: ...Isotemp Plus General Purpose Laboratory Freezers Catalog Numbers 13 986 142 142R 13 986 143 143R 13 986 144 144R 13 986 147 147R 104215 Rev A...

Страница 2: ...Casters 5 Shelves 6 Electrical Connection 6 Operation 7 Control Layout 7 Setting the Temperature Control Setpoint 7 Units Select 8 Calibration Offset 8 Hold Off Time 9 Temperature Digression 9 Automat...

Страница 3: ...s intended application alterations or modifications will void the Warranty This product is not intended nor can it be used as a ster ile or patient connected device In addition this appara tus is not...

Страница 4: ...such external evidence of loss or damage may result in the carrier s refusing to honor your damage claim The carri er will supply the form required to file such a claim Concealed Loss or Damage Conce...

Страница 5: ...it After installation confirm that the unit was installed per these instructions that the required information was completed and that the form was faxed to the specified number Installing and Leveling...

Страница 6: ...e customer should have the wall receptacle and circuit checked by a qualified electrician to verify the receptacle is properly grounded and is connected to the proper power supply The following list g...

Страница 7: ...icated by the C or F LED located just to the right of the temperature display The refrigerator is factory set at 15 C When the unit is first energized there is a delay time before the compressor will...

Страница 8: ...nd control point to a ref erence standard To set the temperature offset press and hold the Menu keypad for 5 seconds The display will flash oS offset followed by the last temperature off set value The...

Страница 9: ...sing the Increase keypad the display will show the warmest temperature recorded since the unit was ener gized or the digression values were reset Pressing the Decrease keypad will display the coldest...

Страница 10: ...t and lukewarm water or a solu tion of bicarbonate of soda 1 tablespoon per gallon of water is recommended for cleaning the interior and exte rior of the cabinet All surfaces should be rinsed and thor...

Страница 11: ...ondensate evaporator pan located behind and below the condenser fan must be cleaned periodically to prevent foul odors and to operate efficiently Vacuum clean if dry or sponge clean with soap and wate...

Страница 12: ...ot A clicking sound occurring about every 30 seconds indicates this protector is working Insufficient cooling Set temp is too high Reduce temperature setting verify Cool LED not lit Cool LED is on Con...

Страница 13: ...t 1 800 438 4851 Item Part No Shelves SPN13986210E Compressor Relay K1 SPN88616 Temperature Controller SPN104161 International Version Controller SPN105402 Chart Recorder SPN104114 Chart Recorder Inst...

Страница 14: ...K DEFROST SWITCH 1 BLK BLK YEL BLU L1 Compressor Relay K1 COMPRESSOR MOTOR CONDENSER FAN S K1 BRN DEFROST TIMER CUT OUT COIL ORN RED DEFROST LIMIT CONTROL WHT DEFROST HEATER CONDENSATE DRAIN HEATER OR...

Страница 15: ...sor Relay Cool Relay R1 CONDENSATE DRAIN HEATER COMPRESSOR MOTOR CONDENSER FAN S DEFROST HEATER WHT BRN PNK R1 DEFROST TIMER CUT OUT COIL DEFROST SWITCH 1 ORN ORN DEFROST LIMIT CONTROL RED BLK YEL YEL...

Страница 16: ...m the respective manufacturer whatever adjustment is reasonable in the light of the manufacturer s own warranty b The Company shall be released from all obligations under its warranty in the event rep...