Restart Procedure
If unit is unplugged or turned off, allow 3 minutes before
restarting or plugging it back in.
Manual Defrost Procedure
(Where Applicable)
Do not use any sharp instrument, blade or scraper to
remove ice and frost on refrigerator surfaces because of
the very real danger of puncturing the cooling coil.
Do not use any electrical device to defrost the unit.
When frost accumulates to 1/4" or more, the operating
efficiency of the unit will be affected. Use the following
procedure to defrost manually:
•
Rotate the temperature control knob to the OFF
setting.
•
Disconnect the power cord from its outlet.
•
Remove contents of unit. If practical, wrap con-
tents in paper and then in a heavy blanket to
maintain temperature of items, especially those
removed from a freezer. Wear protective gloves
to prevent frostbite when handling cold items.
•
Open door and allow free circulation of ambient
air.
•
To speed the process, place pan of warm water
inside the chamber.
•
Wipe out the interior.
•
Replace contents.
•
Reconnect power cord to outlet and set temper-
ature control to desired operating temperature.