Maintenance
Replacing the Door Seal
Fisher Scientific
Peltier Low Temp. Incubators 10-5
Replacing the Door Seal
Special door seals
There are special door seals available, please refer to the
accessory list:
• Viton gasket made for silicone-free applications
• Special gasket for higher humidity applications (reduces risk
of condensation)
The door seal of the outer door is located in the retaining slot.
The door seal should be inspected for any signs of embrittlement at half-yearly intervals.
No tools are required to replace the seal.
Figure 10-1 Door Seal Replacement
1. Pull the seal out of the guide slot.
2. Starting on the hinge side of the door, position the end of the new seal at the location
indicated by the arrow in
above.
3. Gently press the seal into the slot, working around the circumference of the door. In
corner areas in particular ensure that the seal lip is installed without any wrinkles and that
the seal is not stretched or compressed.
4. Make sure that the seal is positioned correctly in the slot and flush with the door frame.
Correct, if necessary.
Содержание Isotemp 178L 120V
Страница 1: ......
Страница 6: ...Contents iv Peltier Low Temp Incubators Fisher Scientific...
Страница 8: ...List of Figures ii Peltier Low Temp Incubators Fisher Scientific...
Страница 26: ...Installation Installing the Anti tilt Anchor 3 10 Peltier Low Temp Incubators Fisher Scientific...
Страница 64: ...Shut down Shutting the Peltier low temperature incubator down 8 2 Peltier Low Temp Incubators Fisher Scientific...
Страница 84: ...Technical Data 13 4 Peltier Low Temp Incubators Fisher Scientific...
Страница 88: ...Device Log 15 2 Peltier Low Temp Incubators Fisher Scientific...
Страница 91: ...Contact Fisher Scientific Peltier Low Temp Incubators 16 3...