Fuse replacement 15 A fuses
The device fuses are not user-serviceable. When the low temperature
incubator exhibits the typical signs of a blown fuse (no response to
pressing the On/Off button, control panel remains extinguished, no
heating operation), call Thermo Fisher Scientific Customer Service to
have the fuses replaced.
Product Description
Data Communications & Alarm Interface
4-10 Peltier Low Temp. Incubators
Fisher Scientific
Replacement 2 A fuses socket
Access to the 2 A fuses for the electrical outlets Fig. 4-10 is performed by turning the grid lock
on the fuse modules at the electronics unit.
Figure 4-10 T 2A H 250V fuse
Fuse replacement 2 A fuses
Only compatible fuses 2 A, slow blow, high breaking current, 250 V are
to be used for replacement.