4.1.2
Clear pH Standardization
The setup option allows you to clear the previous pH standardization.
To clear pH Standardization
1. Access the Clear pH Standardization menu from the pH Measure
screen by pressing the
setup
key.
2. Press
setup
again to display the clear BUFFER icon is displayed
along with the previously entered buffers.
clear
BUFFER
4 7 10
press
to select options
press
to
accept
setup
enter
3. Press
enter
to clear all existing buffers.
4 Press
enter
key to confirm selection OR press
mode
key to exit
to measurement OR press
setup
key to access other setup
options.
NOTE
:
If no buffers have been entered, the meter will skip this
screen.
)
You can escape setup mode at any time by pressing mode.
Pressing enter will always return the display to the Measure
screen after accepting the setup option.
13