4.1.3
pH Buffer Group
This setup option allows you to select from 3 different buffer groups,
for auto buffer recognition.
The 3 buffer groups are:
USA buffers:
2, 4, 7, 10 and 12
European buffers:
1, 3, 6, 8, 10 and 13
NIST buffers:
1.68, 4.01, 6.86, 9.18 and 12.46
To Select pH Buffer Group
1 Access the pH Buffer Group menu from the pH Measure screen
by pressing the
setup
key and the screen will display the
BUFFER select icon.
BUFFER
select
2 4 7 10 12
press
to select options
press
to
accept
setup
enter
2 Press
enter
to accept the group and return to the Measure
screen,
or
3 Continue to press
setup
until the desired buffer group is
displayed.
1 3 6 8 10 13
BUFFER
select
1.68 4.01 6.86 9.18 12.46
BUFFER
select
4 Press
enter
to accept the desired buffer group and return to the
Measure screen.