22
15.
Viewing, Transferring, and Printing Data
To send data directly to a printer or PC you will require the appropriate RS-232 or
USB cable.
Press
VIEW
key to view calibration or stored data directly on the display.
Press
REPORT
soft key to view calibration data directly on the display.
Press
key to print/export stored data manually to a dedicated printer
or computer.
Press
key to print/export calibration data manually to a dedicated
printer or computer.
Use
the
TIMED Datalog
function to print/export live data automatically at
intervals to a dedicated printer or computer (3 second minimum rate). See
EXPORT DATA
within the
SYSTEM SETUP MENU
to use this setting
.
Choose the “
PRINTER
” format option to transfer the data in an easily viewable
format – best for printing.
Choose the “
CSV
” format option to send the data as a comma separated value –
best for exporting data into software
Press
VIEW
and select
DELETE ALL
to erase all saved data that was manually
or automatically stored into memory. This step is most commonly required when
the
TIMED Datalog
interval is not turned off and as a result, fills the memory to
capacity. When the memory is full, deleting the saved data is required in order to
save additional data. The instrument will never erase old data in favor of new data
when the memory is full.
If you wish to view, print or send the saved data to a PC before deleting, use the
PRINT key.
To make changes to the printer, datalog, and output settings, see
Section 3 –
EXPORT DATA.