8
Timed:
When either printing or datalog is ON, the option to select
Timed
readings
ON
or
OFF
appears. Select “
ON
” to automatically print or save data
at selectable intervals. Select “
OFF
” to save data manually using the
PRINT/ENTER
key.
Time Interval:
(MM:SS) (Minutes:Seconds). When timed reading is
ON
, you
must select an interval at which the data will be sent to the Printer and/or Datalog
memory. Choose from 3 seconds to 60 minutes. Data will be collected until the
data limit is reached or the option is turned “OFF”. This feature is useful for
gathering data from a single sample over time. Note that “
SAVE
” is not available
during TIMED setting. To view the stored data, use “
VIEW
” key.
COMM SETUP (Communication Setup)
Format:
To send data as a comma separated value, choose
CSV
– best for
exporting data into spreadsheet software. Choose
PRINTER
to send the data in
an easily viewable format – best for printing.
Communication Setup:
Choose
RS-232
(9600 or 19200 baud rate) or
USB
.
Baud Rate:
When RS-232 is selected, choose (9600 or 19200)
DATE & TIME
Setting the correct date and time is required for GLP and will apply to power on,
measurement, data log, and print functions. Factory reset will not apply to date
and time setting once it has been set. Changes related to daylight savings time
must be manually entered.
Date Format:
Select (MM DD YY) or (DD MM YY)
Time Format:
Select 12Hrs (AM/PM) or 24Hrs
PASSWORD
Select “
ENABLE
” to restrict access to Calibration and Setup modes. When
password protection is enabled, password entry is required before performing any
calibration, or making changes to the setup mode. Setup parameters can be
viewed, but not changed without correct password entry. The password is a user
selectable number from 1 to 99999.
Select “
DISABLE
” if password protection is not desired.
The meter does not allow you to edit setup parameters or perform a new
calibration unless you enter the correct password. If an incorrect password is
entered 3 consecutive times, the meter returns to measurement mode.
If a password has been
ENABLED
but cannot be recovered, a password can be
provided via a written request. The instrument serial number and your contact
information are required.