After installing connecting tubing, check that the PURGE CONTROL valve on the Model 282 is closed (full 
clockwise position) then open the regulator flow control valve to pressurize the line.  At this point, check all 
connections for leaks using an ordinary soapy water solution. 

9.

 

Check data plate and plug line cord into a suitable power receptacle 

OPERATION 

Main Controller Menu 

Energize the oven by switching the ON/OFF toggle to the ON position.  The upper display will show VER while the 
lower display will show 1.0, indicating the current software version.  Next, the DAY/TIME display will begin 
alternating between the day of week and the time of day.  At the same time, the TEMP/VACUUM display will 
alternate between chamber temperature (

°

C) and chamber pressure (inHg).  Press the MENU key to scroll through 

menus 1-6, then press SET to enter a menu.  When there is no keypad activity for 5 seconds, the unit returns to 
menu 1 automatically.  The following table lists the main menus and their functions.  The tables that follow list 
the subcategories of the main menu and describe the function of each. 

 
Table 1: Main Menu 

                                                                                

MAIN MENU 

Menu 

Prompt 

                  Display Functions 

                           Setup Functions 

    1 

(Default) 

Top Display – Day of week, Time of day 
Bottom Display – Temperature (°C), Pressure (in Hg) 

Adjust temperature control setpoint 
Turn vacuum pump contacts On / OFF 

    2 

  ADJT 

Display temperature offset value 

Adjust temperature control offset 

    3 

  ADJP 

Displays pressure offset value 

Adjust for barometric pressure variations 

    4 

  TSET  

Displays flashing day or time value 

Set clock and day of the week 

    5 

 PRGM 

Displays program mode 

Start / Stop program mode 

    6   

  PSET 

Displays day, step and setpoint 

Program 7 – Day event timer 

     

 

Table 2: Default Menu (Menu 1) 

 

 

Step            Key Entry 

                                           Action 

   Display 

   1  

Press SET 

Unit enters 

Temperature Setpoint

 mode with the current setpoint 

displayed.  Use the UP/DOWN keys to adjust the setpoint.   

      SP: 
      280 

   2 

Press SET 2

nd

 Time 

The temperature setpoint is updated and the vacuum pump  
ON/OFF option is displayed.  Use the UP key to switch the 
pump on, and the DOWN key to switch the pump off. 

     PUMP 
      OFF 

   3 

Press SET or wait 2 
seconds 

Unit returns to the default menu. 

     MON   
      280 

 

 

 

 

 

 

    

 

Table 3: Adjust Temperature Offset (Menu 2)  

 

 

Step            Key Entry 

                                           Action 

   Display 

   1  

Press MENU until ADJT 
is displayed, then press 
SET 

Use the UP/DOWN keys to adjust the temperature offset.  
(Range +/- 30 °C) 

     ADJT 
        00 

   2 

Press SET or wait 2 
seconds 

Unit returns to the default menu. 

     MON 
      280 

 

 

 

 

 

 

    

 

Table 4: Adjust Barometric Pressure Offset (Menu 3) 

 

 

Step            Key Entry 

                                           Action 

   Display 

   1  

Press MENU until ADJP 
is displayed, then press 

Use the UP/DOWN keys to adjust the pressure offset.  
(Range +/- 8.0 inHg) 

     ADJP 
       0.0 

Содержание 282A

Страница 1: ...Instructions Fisher Isotemp Model 282A Vacuum Oven Catalog No 13 262 52 53 Manual PN 52063 REV H Published 3 02 2000 Park Lane Pittsburgh PA 15275 412 490 8300...

Страница 2: ...8 Key Entry 8 Table 5 Set Time and Day Menu 4 9 Key Entry 9 Table 6 Program Mode ON OFF Menu 5 9 Key Entry 9 Table 7 Program Edit Mode Menu 6 9 Key Entry 9 Establishing Operating Conditions 10 TEMPERA...

Страница 3: ...s plating and electronic process control Five built in self diagnostic features open thermocouple over temperature power failure low battery and calibration data corruption on the front panel displays...

Страница 4: ...any shipping damage has occurred retain the unit intact including all packing materials and file claim with the final carrier Usually the firm will send an investigator to ascertain liability Table 1...

Страница 5: ...CTORS AND VALVES POWER SWITCH A rocker type switch that controls all power to the Model 282A Vacuum Oven VACUUM CONTROL Valve A forged body shut off valve used to open and close the connection to an a...

Страница 6: ...5 Calibration Jumper Pin JP4 Group of four pins located in the center of the rear of the control board The lower two pins when jumped enable calibration mode See Figure 2 for controller rear view...

Страница 7: ...point 20 C PWRF Power Failure Power failure only detected during program mode The failure must be longer than 2 minutes Time display alternates between PWRF HH MM and DAY n at 5 second intervals BATT...

Страница 8: ...m oven for operation 1 Verify that all packing items and securing materials have been removed from the oven 2 Move the oven to the desired location as specified above 3 Locate the two serrated hose fi...

Страница 9: ...Time of day Bottom Display Temperature C Pressure in Hg Adjust temperature control setpoint Turn vacuum pump contacts On OFF 2 ADJT Display temperature offset value Adjust temperature control offset 3...

Страница 10: ...Program Edit Mode Menu 6 Step Key Entry Action Display 1 Press MENU until PSET is displayed then press SET Unit enters the program edit mode and points to MONday step 1 Use the UP DOWN keys to select...

Страница 11: ...Tabs 3 and 4 NOTE Vacuum pump operation can be performed independently as well 1 Remove power cord from the supply outlet 2 To access the auxiliary contacts remove six screws holding the right side p...

Страница 12: ...er the value Return to the default mode by continuously pressing MENU or leave as is for two seconds and control will revert to the default mode automatically NOTE If fast warm up is desired the tempe...

Страница 13: ...the UP DOWN arrow keys Press the SET key again to enter the day of the week PROGRAMMING THE OVEN The oven controller is capable of handling 8 instructions per day 7 days a week Programming the control...

Страница 14: ...Pump OFF Time 1 00 PM Temp 150 C Pump OFF Time 1 00 PM Temp OFF Pump ON Time 12 00 AM Temp OFF Pump ON Time 12 00 AM 4 Temp OFF Pump OFF Time 3 00 PM Temp OFF Pump OFF Time 1 00 PM Temp OFF Pump OFF T...

Страница 15: ...tude and atmospheric conditions Calibration Mode Follow the instructions below to access the calibration mode 1 Set the POWER switch to the OFF position 2 Remove the right side panel from the Model 28...

Страница 16: ...ry that the barometric or atmospheric pressure be known One way of doing this is to contact the U S Weather Bureau in the installation area The bureau will provide the barometric pressure corrected to...

Страница 17: ...ge McLeod and convert the reading from millimeters to inches of mercury divide by 25 4 mm inches 10 Using the UP DOWN arrow keys adjust the vacuum display to match the reading converted from the refer...

Страница 18: ...DOWN arrow keys to increase or decrease the vacuum output voltage then press the SET key to enter the new value The output voltage will be updated Repeat the process until the vacuum output voltage i...

Страница 19: ...oth left and right sides of the door using your hands Close vacuum valve at 25 inches of Hg 7 Tighten the two bolts for each hinge while lightly pressing each corner of the door in the area of the hin...

Страница 20: ...19 Figure 6 1 6 6 Door Alignment...

Страница 21: ......

Страница 22: ...alysis Chart Symptom Probable Cause Corrective Action Oven does not heat CRC error Error found in calibration data Recalibrate unit Oven does not heat OTC error Open Control Thermocouple Check T C con...

Страница 23: ...e publication date see front cover of this manual and subsequent revisions may have occurred after publication Table 11 Replacement Parts DESCRIPTION PART NUMBER Control P C Board 115 230V SPN 103305...

Страница 24: ...23...

Страница 25: ...24 Schematics Figure 7 115V Schematic Figure 8 230V Schematic...

Отзывы: