Temperature/Pressure Offsets 

When the oven reaches a stable operating condition, the display temperature should indicate the actual center 
chamber temperature (+/- 2.0°C).  In the event a more accurate display reading is needed, the offset parameter, 
located in the MAIN MENU, can compensate for the error.  To enter an offset value, press the MENU key until the 
upper display window shows ADJT (adjust temperature) then press UP/DOWN arrow keys to enter an offset for the 
display temperature.  Press the MENU key 5 times to return to the default display or wait for 2 seconds and 
control will shift automatically.  
 
EXAMPLE:  The oven is temperature is set for 200°C.  After it has reached a stable state (usually 2½ hours for 
200°C) the actual temperature of the oven center, as read by a certified thermometer, is 203°C.  The oven display 
shows 200°C.  Entering a +3 for the ADJT parameter will move the display to agree with the actual oven 
temperature.   
 
NOTE:  Adjusting the offset only changes the display temperature.  It does not alter the controller operation. 
 
The vacuum display can be adjusted in the same way.  In the event the displayed chamber pressure is not 
accurate, press the MENU key until the upper display window shows ADJP (adjust pressure).  Use the UP/DOWN 
keys to enter the desired offset.  Press the MENU key four times to return to the default display or wait for 2 
seconds and control will shift automatically.  The displayed chamber pressure will be updated with the offset 
value. 
 

Setting the Time and Day 

To set the time and day press the MENU key until the upper display shows TSET (time set).  Press the SET key 
and use the UP/DOWN arrow keys to adjust the hour of the day.  When the current hour is displayed enter it by 
pressing the SET key again.  Use the UP/DOWN arrow keys to adjust the minutes.  When the correct minutes are 
displayed, press the SET key.  Finally, days of the week can also be adjusted by pressing the UP/DOWN arrow 
keys.  Press the SET key again to enter the day of the week.   

 
PROGRAMMING THE OVEN 

The oven controller is capable of handling 8 instructions per day, 7 days a week.  Programming the controller is 
not difficult.  Table 7, listed as 

MENU 6 – Program Edit Mode

, provides step by step instructions for setting the 

control temperature and/or vacuum pump operation at any given time of day or day of the week.  Use Table 7 in 
conjunction with the following description when programming the oven controller 
   
To enter the 

Program Edit Mode

 press the MENU key until the upper display shows PSET.  The upper display will 

then show MON1, indicating the first step in Monday.  Use the UP/DOWN arrow keys to step through to the 
current day of the week.  When the desired step is reached press SET.  Use the UP/DOWN arrow keys to adjust 
the starting hour of the current step, press SET to select the desired hour.  Repeat for minutes, temperature and 
vacuum pump operation.  The next prompt (N ?) asks the user to copy the current step to the remaining steps in 
the day.  Use the UP arrow key to select Y (yes), or leave as is and select no by pressing the SET key.  The upper 
display will then shift to the next step, MON2 (if MON1 was the previous step).  Repeat this procedure for entering 
new steps.   
If a step was copied to the end of the day, that same day can also be copied to the end of the week.  To do this, 
press the MENU key when the next step appears in the upper display window.  At that point, press the MENU key.  
The upper display will show COPY, while the lower display shows NO.  Use the UP arrow key to change the NO to 
YES.  Press the SET key and the previous day’s program is copied to the end of the week. 
 
NOTE:  The program week begins on Monday and ends on Sunday.     

 
Sample Program: 

 
Monday: 
1.

 

Evacuate chamber and operate at 100°C for 5 hours starting at 8AM.  Maintain less than 5 in Hg during the 5 
hours. 

2.

 

Increase temperature to 150°C and operate for 2 hours.  Maintain same chamber pressure. 

3.

 

Shut down oven for remainder of day. 

Tuesday: 
 

Repeat steps from Monday except step 2.  Have oven shut down after 100°C operation.   

Wednesday: 

Содержание 282A

Страница 1: ...Instructions Fisher Isotemp Model 282A Vacuum Oven Catalog No 13 262 52 53 Manual PN 52063 REV H Published 3 02 2000 Park Lane Pittsburgh PA 15275 412 490 8300...

Страница 2: ...8 Key Entry 8 Table 5 Set Time and Day Menu 4 9 Key Entry 9 Table 6 Program Mode ON OFF Menu 5 9 Key Entry 9 Table 7 Program Edit Mode Menu 6 9 Key Entry 9 Establishing Operating Conditions 10 TEMPERA...

Страница 3: ...s plating and electronic process control Five built in self diagnostic features open thermocouple over temperature power failure low battery and calibration data corruption on the front panel displays...

Страница 4: ...any shipping damage has occurred retain the unit intact including all packing materials and file claim with the final carrier Usually the firm will send an investigator to ascertain liability Table 1...

Страница 5: ...CTORS AND VALVES POWER SWITCH A rocker type switch that controls all power to the Model 282A Vacuum Oven VACUUM CONTROL Valve A forged body shut off valve used to open and close the connection to an a...

Страница 6: ...5 Calibration Jumper Pin JP4 Group of four pins located in the center of the rear of the control board The lower two pins when jumped enable calibration mode See Figure 2 for controller rear view...

Страница 7: ...point 20 C PWRF Power Failure Power failure only detected during program mode The failure must be longer than 2 minutes Time display alternates between PWRF HH MM and DAY n at 5 second intervals BATT...

Страница 8: ...m oven for operation 1 Verify that all packing items and securing materials have been removed from the oven 2 Move the oven to the desired location as specified above 3 Locate the two serrated hose fi...

Страница 9: ...Time of day Bottom Display Temperature C Pressure in Hg Adjust temperature control setpoint Turn vacuum pump contacts On OFF 2 ADJT Display temperature offset value Adjust temperature control offset 3...

Страница 10: ...Program Edit Mode Menu 6 Step Key Entry Action Display 1 Press MENU until PSET is displayed then press SET Unit enters the program edit mode and points to MONday step 1 Use the UP DOWN keys to select...

Страница 11: ...Tabs 3 and 4 NOTE Vacuum pump operation can be performed independently as well 1 Remove power cord from the supply outlet 2 To access the auxiliary contacts remove six screws holding the right side p...

Страница 12: ...er the value Return to the default mode by continuously pressing MENU or leave as is for two seconds and control will revert to the default mode automatically NOTE If fast warm up is desired the tempe...

Страница 13: ...the UP DOWN arrow keys Press the SET key again to enter the day of the week PROGRAMMING THE OVEN The oven controller is capable of handling 8 instructions per day 7 days a week Programming the control...

Страница 14: ...Pump OFF Time 1 00 PM Temp 150 C Pump OFF Time 1 00 PM Temp OFF Pump ON Time 12 00 AM Temp OFF Pump ON Time 12 00 AM 4 Temp OFF Pump OFF Time 3 00 PM Temp OFF Pump OFF Time 1 00 PM Temp OFF Pump OFF T...

Страница 15: ...tude and atmospheric conditions Calibration Mode Follow the instructions below to access the calibration mode 1 Set the POWER switch to the OFF position 2 Remove the right side panel from the Model 28...

Страница 16: ...ry that the barometric or atmospheric pressure be known One way of doing this is to contact the U S Weather Bureau in the installation area The bureau will provide the barometric pressure corrected to...

Страница 17: ...ge McLeod and convert the reading from millimeters to inches of mercury divide by 25 4 mm inches 10 Using the UP DOWN arrow keys adjust the vacuum display to match the reading converted from the refer...

Страница 18: ...DOWN arrow keys to increase or decrease the vacuum output voltage then press the SET key to enter the new value The output voltage will be updated Repeat the process until the vacuum output voltage i...

Страница 19: ...oth left and right sides of the door using your hands Close vacuum valve at 25 inches of Hg 7 Tighten the two bolts for each hinge while lightly pressing each corner of the door in the area of the hin...

Страница 20: ...19 Figure 6 1 6 6 Door Alignment...

Страница 21: ......

Страница 22: ...alysis Chart Symptom Probable Cause Corrective Action Oven does not heat CRC error Error found in calibration data Recalibrate unit Oven does not heat OTC error Open Control Thermocouple Check T C con...

Страница 23: ...e publication date see front cover of this manual and subsequent revisions may have occurred after publication Table 11 Replacement Parts DESCRIPTION PART NUMBER Control P C Board 115 230V SPN 103305...

Страница 24: ...23...

Страница 25: ...24 Schematics Figure 7 115V Schematic Figure 8 230V Schematic...

Отзывы: