THE INSTRUMENT 

The Fisher Isotemp® Vacuum Oven, Model 282A, is designed to operate with reduced pressures and/or with inert 
atmospheres.  As such, it is equipped with separate vacuum and gas ports--each has its own control valve.  
Additionally, a special digital readout display indicates time of day, chamber temperature, pressure and schedules 
for programmed oven operation. 
Temperature settings and vacuum pump operation can be controlled at the touch of a button.  A 7-day 
programmable timer provides automatic temperature/vacuum control in real time programmable steps.  The dual 
displays provide constant day/time and temperature/pressure updates every 5 seconds.  Chamber temperature 
and pressure can be externally recorded by connecting to the recorder outputs located on the back of the control 
board.  PURGE and VACUUM valves, conveniently located under the control panel, provide access for a vacuum 
pump and/or the presence of inert gases into the chamber. 
The Model 282A design features a 1.5 cubic-foot stainless steel chamber with fully adjustable shelves.  An 
additional feature is a chamber door with full view, and tempered safety glass. 
The stainless steel chamber permits performance of practically every laboratory need associated with general 
drying, curing, conditioning, desiccating, annealing, and moisture tests.  Additional applications include vacuum 
embedding, out-gassing solids and liquids, aging tests, plating, and electronic process control. 
Five built-in, self diagnostic features--open thermocouple, over temperature, power failure, low battery, and 
calibration data corruption--on the front panel displays to alert the operator of a malfunction and thus prevent 
damage to both samples and equipment. 

Performance Characteristics 

Temperature 

Temperature Range .................................... Ambient to 280ºC continuously adjustable with 1ºC resolution 
Heat Rise time............................................ Ambient to 100ºC in <30 minutes, with temp set to 200°C. 
Temperature Set-Point Accuracy ................. <±2º C from 50 to 280°C  
Cool-Down Time .......................................... 280 to 100ºC in approximately 2.5 hours 
Temperature Set-Point Reproducibility......... ±0.1ºC 
Temperature Control Band .......................... ±0.2ºC 
Long-Term (12 hours) Temp. Stability........... ±0.03ºC 
Temperature Display................................... 3 digit LED display to nearest to nearest 1ºC 
Average Temperature Uniformity……….………<±5ºC (in vacuum),based upon measuring points at the 
 

geometric center and each corner of oven chamber at 100

°

Vacuum 

Vacuum Pump Down Time ........................... To <1 torr in 3 minutes using a pump having a free air 
 

displacement of 100 LPM; 8 minutes using a 35 LPM pump 

Average Leak Rate ...................................... <1 inch Hg per 24 hours 
Maximum Vacuum ...................................... Evacuates to <1 torr (0.039 inch Hg) absolute 
Vacuum Display Accuracy ........................... Average error 0.14 inch Hg between ambient and complete 
 

vacuum 

Atmosphere Control .................................... Use with non-corrosive, non-flammable gases such as nitrogen 
 

and CO

2

 

Power Requirements 

Catalog No. 13-262-52 ................................. 115V ±10%, 14.7 Amps, 50/60 Hz 
Catalog No. 13-262-53 ................................. 230V ±10%, 7.4 Amps, 50/60 Hz 
Power Consumption..................................... 1700 watts 
Vacuum Pump Contacts (Tabs 3&4).............. 1 HP @ 115 VAC, 2 HP @ 230VAC 

 

Environmental

Operating Temperature ............................... 60º to 100ºF 
Storage Temperature .................................. 32º to 120ºF 

Содержание 282A

Страница 1: ...Instructions Fisher Isotemp Model 282A Vacuum Oven Catalog No 13 262 52 53 Manual PN 52063 REV H Published 3 02 2000 Park Lane Pittsburgh PA 15275 412 490 8300...

Страница 2: ...8 Key Entry 8 Table 5 Set Time and Day Menu 4 9 Key Entry 9 Table 6 Program Mode ON OFF Menu 5 9 Key Entry 9 Table 7 Program Edit Mode Menu 6 9 Key Entry 9 Establishing Operating Conditions 10 TEMPERA...

Страница 3: ...s plating and electronic process control Five built in self diagnostic features open thermocouple over temperature power failure low battery and calibration data corruption on the front panel displays...

Страница 4: ...any shipping damage has occurred retain the unit intact including all packing materials and file claim with the final carrier Usually the firm will send an investigator to ascertain liability Table 1...

Страница 5: ...CTORS AND VALVES POWER SWITCH A rocker type switch that controls all power to the Model 282A Vacuum Oven VACUUM CONTROL Valve A forged body shut off valve used to open and close the connection to an a...

Страница 6: ...5 Calibration Jumper Pin JP4 Group of four pins located in the center of the rear of the control board The lower two pins when jumped enable calibration mode See Figure 2 for controller rear view...

Страница 7: ...point 20 C PWRF Power Failure Power failure only detected during program mode The failure must be longer than 2 minutes Time display alternates between PWRF HH MM and DAY n at 5 second intervals BATT...

Страница 8: ...m oven for operation 1 Verify that all packing items and securing materials have been removed from the oven 2 Move the oven to the desired location as specified above 3 Locate the two serrated hose fi...

Страница 9: ...Time of day Bottom Display Temperature C Pressure in Hg Adjust temperature control setpoint Turn vacuum pump contacts On OFF 2 ADJT Display temperature offset value Adjust temperature control offset 3...

Страница 10: ...Program Edit Mode Menu 6 Step Key Entry Action Display 1 Press MENU until PSET is displayed then press SET Unit enters the program edit mode and points to MONday step 1 Use the UP DOWN keys to select...

Страница 11: ...Tabs 3 and 4 NOTE Vacuum pump operation can be performed independently as well 1 Remove power cord from the supply outlet 2 To access the auxiliary contacts remove six screws holding the right side p...

Страница 12: ...er the value Return to the default mode by continuously pressing MENU or leave as is for two seconds and control will revert to the default mode automatically NOTE If fast warm up is desired the tempe...

Страница 13: ...the UP DOWN arrow keys Press the SET key again to enter the day of the week PROGRAMMING THE OVEN The oven controller is capable of handling 8 instructions per day 7 days a week Programming the control...

Страница 14: ...Pump OFF Time 1 00 PM Temp 150 C Pump OFF Time 1 00 PM Temp OFF Pump ON Time 12 00 AM Temp OFF Pump ON Time 12 00 AM 4 Temp OFF Pump OFF Time 3 00 PM Temp OFF Pump OFF Time 1 00 PM Temp OFF Pump OFF T...

Страница 15: ...tude and atmospheric conditions Calibration Mode Follow the instructions below to access the calibration mode 1 Set the POWER switch to the OFF position 2 Remove the right side panel from the Model 28...

Страница 16: ...ry that the barometric or atmospheric pressure be known One way of doing this is to contact the U S Weather Bureau in the installation area The bureau will provide the barometric pressure corrected to...

Страница 17: ...ge McLeod and convert the reading from millimeters to inches of mercury divide by 25 4 mm inches 10 Using the UP DOWN arrow keys adjust the vacuum display to match the reading converted from the refer...

Страница 18: ...DOWN arrow keys to increase or decrease the vacuum output voltage then press the SET key to enter the new value The output voltage will be updated Repeat the process until the vacuum output voltage i...

Страница 19: ...oth left and right sides of the door using your hands Close vacuum valve at 25 inches of Hg 7 Tighten the two bolts for each hinge while lightly pressing each corner of the door in the area of the hin...

Страница 20: ...19 Figure 6 1 6 6 Door Alignment...

Страница 21: ......

Страница 22: ...alysis Chart Symptom Probable Cause Corrective Action Oven does not heat CRC error Error found in calibration data Recalibrate unit Oven does not heat OTC error Open Control Thermocouple Check T C con...

Страница 23: ...e publication date see front cover of this manual and subsequent revisions may have occurred after publication Table 11 Replacement Parts DESCRIPTION PART NUMBER Control P C Board 115 230V SPN 103305...

Страница 24: ...23...

Страница 25: ...24 Schematics Figure 7 115V Schematic Figure 8 230V Schematic...