4.2 Speed and Force Tables
Rotor 04-976-008 4-Place Swinging Bucket, 4x750 ml
Complete with 4 buckets and 1 mounted rotor recognition disk, but without Sealed tube adapters
Tube Maximum
Adapter*
No. x Vol. (ml)
Description
RPM / RCF Radius Quantity x Cat. No.
4 x 750ml unsealed
05-112-76 bottle
3400
2550
19.6
Not required
4 x 500ml unsealed
Use 05-112-76 bottle
(graduated to 600ml)
3400
2550
19.6 Not
required
4 x 250ml unsealed
Corning conical 05-538-53 bottle
3400
2625
20.3
4 x 04-974–008K
4 x 250ml unsealed
Corning round bottom glass 05-586
Corning flat bottom glass 05-587A
3400
2625
18.1
4 x 04-974-008B
4 x 250ml sealed
05-432A, 05-432B bottles
3400
2340
18.1
4 x 04-974-008B
4 x 175/225ml sealed
Falcon conical
05-538-62/05-538-61
3400
2340
18.1
4 x 04-974-008C
8 x 100ml sealed
38 x 111 mm
3400
2310
17.9
4 x 04-974-002S
20 x 50ml sealed
Falcon/Corning conical
3400
2370
18.3
4 x 04-974-005F
28 x 50ml sealed
29x116mm round bottom glass/plastic 3400
2480
19.2
4 x 04-974-008E
28 x 25-30ml sealed
25 x 105 mm
3400
2380
18.4 Use 7x05-375-11 in
each 04-974-008E
48 x 15ml sealed
Falcon/Corning conical
3400
2440
18.9
4 x 04-974-008G
48 x 10ml sealed
Kova/Urisystem
3400
2440
18.9
4 x 04-974-008G
48 x 10-15ml sealed
Vacutainer 16x100, 16x125mm
3400
2520
19.5
4 x 04-974-008L
60 x 7-10ml sealed
Vacutainer 16x75-100mm
3400
2260
17.5
4 x 04-974-008H
100 x 7ml sealed
Vacutainer 13x100mm
3400
2220
17.2
4 x 04-974-008I
100 x 7ml sealed
Hemogard Vacutainer 13x100mm
3400
2220
17.2
4 x 04-974-008I
100 x 5ml sealed
Vacutainer 13x75mm
3400
2220
17.2
4 x 04-974-008I
100 x 5ml sealed
Hemogard Vacutainer 13x75mm
3400
2220
17.2
4 x 04-974-008I
148 x 3-5ml sealed
10x75, 12x 75mm, Falcon # 2063
3400
2050
15.9
4 x 04-974-008J
100 x 1.5/2ml sealed
microtubes
3400
2050
15.9
4 x 04-974-008I
100 x 0.7ml sealed
Microtainers
3400
2050
15.9
4 x 04-974-008I
& 9 pk/12 05-152-3
100 x 0.5ml sealed
PCR microtubes
3400
2050
15.9
4 x 04-974-008I
& 9 pk/12 05-152-3
100 x 0.4ml sealed
microtubes
3400
2050
15.9
4 x 04-974-008I
& 9 pk/12 05-152-4
100 x 0.25ml sealed
microtubes
3400
1930
14.9
4 x 04-974-008I
& 9 pk/12 05-152-4
* Adapters with Number Series 04-977-00xx are sold by the pack/2.
Note: Sealed Adapters comply with the aerosol containment requirements of:
OSHA 29 CFR Part 1910.1030 Occupational Exposure to Bloodborne Pathogens; Final Rule dated
12/06/91
Falcon, Vacutainer and Microtainer are registered trademarks of Becton Dickinson & Co.
MARATHON Series Operation Manual
28
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com